Chapter 4 - Configuration
4-17
Some Things to
Remember When
Changing the
Jumper Settings
•
The number of averages selected in the Tuning/Control menu adds
to the read time.
•
For example: With the jumper set in the W4 position and the
number of averages set at 20, the final weight reading would not
be reached until 2 seconds (1 second for the WAVERSAVER
and 1 second for 20 averages at 20 updates per second).
•
See Table 4-7 when changing jumpers for the setting best suited for
your application requirements.
TABLE 4-7: JUMPER SETTINGS AND RESPONSE TIME
JUMPER
MINIMUM
FREQUENCY
APPROXIMATE RESPONSE TIME
W1
7.5 Hz
Least vibration immunity. Fastest time to
reach the final weight reading (approximately
20mS).
W2
3.5 Hz
100 mS to reach the final weight reading.
W3
1.0 Hz
500 mS to reach the final weight reading.
W4
0.5 Hz
Default setting from the factory. Very good
vibration immunity requiring approximately 1
second to reach the final weight reading.
W5
0.25 Hz
Best vibration immunity requiring approxi-
mately 2 seconds to reach the final weight
reading.