TROUBLE
PROBABLE CAUSE
REMEDY
1.
Loss of Memory
2.
Controller locks up
or very erratic con-
troller performance
Frequent Power
brown outs/black
outs
1. Restore data from the
Secure Memory Mod-
ule. (See Chapter 3 for
installation instruc-
tions) (See Chapter 5
for Memory Restore
instructions.
2. Install a power surge
suppressor. (See
Chapter 6 for instruc-
tions)
Chapter 9 - Torubleshooting and Repair
9-7
•
Weight Readings are Unstable, Modulate Excessively
TROUBLE
PROBABLE CAUSE
REMEDY
Weights continuously
modulate up and down.
Excessive vibration
in the process facility.
Set WAVERSAVER to W4 or
adjust until you reach stable
weight readings (See Chap-
ter 3 for instructions)
Weights Displayed are
Incorrect
After Calibration
Display Weight
Reading is Incorrect
TROUBLE
PROBABLE CAUSE
REMEDY
The Weight Display indi-
cates zero with a weight
on the scale.
The Weight Indication
does not change.
Input from the Load
Cell is negative. Load
cell correct input
should always be
between 0 and +30
mV DC.
Check the current counts.
Measure pins 3 & 4 for the
millivolt load cell output.
•
Checking the Current Count
Step 1. Push the Test Button on the Front Panel. The Test Sub-menu
appears.
Step 2. Press the down arrow
↓
until the cursor is to the left of the
current count (Current Ct.) sub-menu. (See Fig. 9-3)
Result 1. If the Current Count is correct, contact your local representa-
tive or
Hardy Process Solutions
for assistance.
Summary of Contents for HI 2160RC PLUS
Page 1: ...HI 2160RCPLUS Rate Controller Operation and Installation Manual ...
Page 23: ...HI 2160RCPLUS MANUAL 2 6 ...
Page 49: ...HI 2160RCPLUS MANUAL 3 26 ...
Page 67: ...HI 2160RCPLUS MANUAL 4 18 ...
Page 91: ...HI 2160RCPLUS MANUAL 5 24 ...
Page 151: ...HI 2160RCPLUS MANUAL 7 12 ...
Page 187: ...HI 2160RCPLUS MANUAL 8 36 ...
Page 207: ...HI 2160RCPLUS MANUAL 9 20 ...
Page 219: ...Data Worksheets WKSHT 6 ...