Hantek HDP135V6 User Manual Download Page 3

 

 

VI. Safety Instructions 

1 Check the AC Voltage Gear 

The HDP135V6 power supply supports two 

specifications of AC powerinputs(115VAC  or 

230VAC).Before use, the user should check the gear 

of the AC voltage selector on the rear panel of the 

power supply, and the input voltage should be within 

the range allowed by the gear (±10%). 

2 Check the Fuse 

The instrument has been installed with fuse of 

specified specification at delivery.Before use, please 

check whether the fuse model matches the AC 

voltage gear.If the fuse does not match or is blown, 

replace the fuse according to the specification. 

Procedure for replacing a fuse: 

1. Turn off the power supply and remove the power 

cable. 

2. Take out the fuse holder. 

 

3. Remove the fuse and install a new one. 

4. Take the fuse holder back into the slot. 

3 Connect the AC Power Supply 

Use the power cable provided in the accessories to 

connect the instrument to the properly grounded AC 

power supply. 

To avoid electric shock, please make sure that the 

instrument is properly grounded. 

4 Startup 

Press the power switch to start the instrument. The 

display screen will be lightened up. 

VII. RS232 Remote Control   

Only HDP135V6S supports RS232 communication. 

To use RS232 for remote control, please log in to the 

following website to download: 

http://hantek.com.cn/products/detail/12272

 

Or scan the QR code below to enter the website: 

 

VIII. Troubleshooting 

1 If the AC  power supply voltage is normal and the 

display screen does not light after the startup, it may be 

a fuse blown or other faults.Turn off the instrument, 

disconnect the power cable, replace the fuse, or ask a 

professional for inspection and repair. 

2 For constant voltage output, if the output voltage is 

less than the setting value and the CC light is on, which 

indicates that the instrument  is in  current protection. 

Itwill  automatically switch to constant current output 

state. At this time,  check the load or increase the 

maximum current according to the use situation. 

3  For  constant current output, if the output  current is 

less than the setting value and the CV light is on, which 

indicates that the instrument is in open circuit voltage 

protection.  It  will  automatically switch to constant 

voltage  output  state. At this time,  check the load or 

increase the maximum voltage according to the use 

situation. 

4 If the fault persists, please contact Hantek. 

IX.Packing List 

1 One Power Supply 

2 One Power Cable 

3 One AlligatorClip Test Line 

4 Certificate of Conformity 

 