HDP135V6 DC Power Supply User Manual
I. Product Introduction
HDP135V6 power supply is a high quality,
cost-effectiveDC power supply. It has four-digit LED
voltage and current display. It also has over voltage
protection (OVP), over current protection (OCP), over
temperature protection (OTP) and USB quick charging
functions.
II. Product Model
Model
CH
Output V/A
LCD
digits
RS232
AC input
HDP135V6
1
35.2V/6.2A
4
None
115/230VAC
HDP135V6A
1
35.2V/6.2A
4
None
230VAC
HDP135V6B
1
35.2V/6.2A
4
None
115VAC
HDP135V6S
1
35.2V/6.2A
4
Yes
115/230VAC
III. PerformanceSpecifications
1 Rated Working Condition
AC input: 115 VAC ± 10%, 230 VAC ± 10%, 50/60Hz
Working Conditions: Temperature -10
℃
~40
℃
R
elative humidity ≤80%
Storage Conditions: Temperature -20
℃
~60
℃
R
elative humidity ≤80%
2 Specifications
All technical specifications can only be guaranteed
when the instrument is operated continuously for more
than 30 minutes at the specified operating temperature
(18
℃
to 28
℃
).
Voltage
Load Regulation
≤0.1% + 5mV
Line Regulation
≤0.01% + 5mV
Set Resolution
10mV
Set Accuracy
≤0.1% + 1 digits
Readback
Resolution
10mV
Readback Accuracy
≤0.1% + 1 digits
Ripple
10mVrms
Current
Load Regulation
≤0.2% + 3mA
Line Regulation
≤0.2% + 3mA
Set Resolution
1mA
Set Accuracy
≤0.2% + 3mA
Readback
Resolution
1mA
Readback Accuracy
≤0.2% + 3 digits
Ripple
5mArms
OVP
0 ~ 35.2V ± 0.2%FS
Maximum Voltage
35.2V ± 0.2%
OCP
0 ~ 6.2A ± 0.2%FS
Maximum Current
6.2A ± 0.2%
Quick Charge
Output Voltage
Range
5V ~ 12V
Output Power
Maximum: 18W
Supported Quick
Charging Protocol
DCP (Apple, Samsung and BC1.2),
Qualcomm QC2.0/QC3.0, Huawei
Quick Charge Protocol FCP, Samsung
Quick Charge Protocol AFC
General Features
Remote Interface
RS232 (Optional)
Baud rate: 2400
Data bit: 8
Stop bit: 1
No parity check
AC Input
115 VAC ± 10%, 230 VAC ± 10%,
50/60Hz
Working Conditions
Temperature -10
℃
to 40
℃
Relative Humidity ≤ 80%
Storage Conditions
Temperature -20
℃
to 60
℃
Relative Humidity ≤ 80%
Cooling
Air cooling
Weight
2.2 Kg
Size
85 * 150 * 230 (mm)
IV.Front Panel
1 Voltage Display
2 Current Display
3 Save and Recall Keys M1 ~ M10
M1 ~ M4: Press
[M*]
once, and the indicator lights
up to display the group settings.Use the voltage and
current adjustment knobs to adjust the settings that
need to be saved, and then long press the
[M*]
key
until the indicator lights up, the settings will be saved
in the
[M*]
position.
M5 ~ M10: To use M5 ~ M10 to save and recall, use
it in conjunction with the
[Shift]
key.Taking
M5
as an
example, press the
[Shift]
key and the
[Shift]
1