Basic Waveform Output
EN
Copyright © Qingdao Hantek Electronics Co., LTD HDG3000B User manual
19
display prompt message and set the frequency to the frequency upper limit of the new
function automatically.
Press
[Setting] >
Frequency
softkey. At this point, use the numeric keypad to input the
value of the frequency and select the desired unit from the pop-up unit menu, or use the
direction keys and knob to modify the current value.
1.
For the input method of frequency value, please refer to the introduction of
".
2.
The frequency units available are MHz, KHz, Hz and mHz.
3.
Press
[Utility] >
Units
softkey and press Frequency/Period softkey to switch
Frequency in the parameter to Period. Press
[Setting] >
Period
softkey to modify
the period parameter.
4.
The period units available are sec, msec, μsec and nsec.
5.2.3 Set the Amplitude
The amplitude range is limited by Frequency or Period and Impedance settings. Please
refer to "Output Characteristics" in Specifications. The default value is 200mVpp.
The amplitude displayed on the screen is the default value or the amplitude previously
set. When the instrument configuration (e.g. frequency) is changed, if this amplitude is
valid, the instrument will still use this amplitude; otherwise, the instrument would display
a prompt message and set the amplitude to the amplitude upper limit of the new
configuration automatically.
Press
[Setting] >
Amplitude
softkey. At this point, use the numeric keypad to input the
value of the amplitude and select the desired unit from the pop-up unit menu, or use the
direction keys and knob to modify the current value.
1.
For the input method of amplitude value, refer to the introduction in "
".
2.
The amplitude units available are Vpp, mVpp, and Vrms, dBm (50Ω impedance).
3.
Press
[Utility] >
Units
softkey and press
Amplitude
/
High
Level
softkey to switch
the
Amplitude
and
Offset
in the parameter to
High Level
and
Low level
. Press
[Setting] >
High Level
or
Low Level
softkey to modify high level or low level
parameters.
4.
The high level units available are V and mV.
Tips:
Switch between Vpp and Vrms
Vpp is the unit for the signal peak-peak value and Vrms is the unit for the signal effective
value. The default unit of the instrument is Vpp. Users can quickly switch the current
amplitude unit from the front panel.
For different waveforms, the relations between Vpp and Vrms are different. Take sine
waveform as an example; the relation of the two units is as shown in the figure below.