Basic Operation
DSO1000B Series HandHeld Oscilloscope User Manual 31
Rise Time
Measure the time between 10% and 90% of the first rising
edge of the waveform.
Fall Time
Measure the time between 90% and 10% of the first falling
edge of the waveform.
Positive Width
Measure the time between the first rising edge and the next
falling edge at the waveform 50% level.
Negative Width
Measure the time between the first falling edge and the next
rising edge at the waveform 50% level.
Off
Do not take any measurement.
Taking Measurements:
For a single waveform (or a waveform divided among multiple
waveforms), up to 8 automatic measurements can be displayed at a time. The waveform channel
must stay in an ‘ON’ (displayed) state to facilitate the measurement. The automatic measurement
can not be performed on reference or math waveforms, or in XY or Scan mode.
5.5.3 ACQUIRE
Push the ACQUIRE button to set the acquisition parameter.
Options
Settings
Comments
Category
Real Time
Equ-Time
Acquire waveforms by real-time digital technique.
Rebuild waveforms by equivalent sample technique.
Mode
(Real Time)
Normal
Peak Detect
Average
Acquire and accurately display most waveforms.
Detect glitches and eliminate the possibility of
aliasing.
Reduce random or uncorrelated noise in signal
display. The number of averages is selectable.
Averages
(Real Time)
4
16
64
128
Select the number of averages.
The readouts in big font size on
the menu are just results of the
corresponding measurements.