C. Installation
Phase One Disintegration Tester User Guide—40-108-001 Rev. A—14 November 2016
C-2
Unpack the Tester
While unpacking the tester, first make note of any damage to
the shipping container. If the container is damaged, contact the
shipper immediately.
1. Remove straps from shipping container.
2. Slide box up and away from enclosed components.
3. Remove any accessory boxes.
4. Carefully remove tester from shipping container.
5. Remove blue tape securing arms and remove plastic
wrap covering the bath lid.
6. Remove the bath lid.
7. Clean waterbath of packing material or other debris.
8. Place tester in final location on bench.
Assemble the Tester
1. Place heater into bath compartment (do not secure).
2. Install bath lid on top of bath while centering the heater
in the slot in the bath lid.
3. Clamp heater into place using knob on back of heater.
4. Use the cable clamps provided with the tester to route
the heater power cable, if desired.
5. Note: beakers and baskets cannot be installed until the
tester is initialized (Section D – Operation).
Electrical Connections
1. Ensure power switch on left back side of tester is in the
Off (O) position.
2. Ensure power switch back of heater unit is in the Off
(O) position.
3. Insert power cable into back of tester; plug outlet end
into power source.
4. Plug heater power cord into power source outlet.