Fitting guide.

Slim Tube HE for BTE-products.

Positioning the hearing instrument on the ear.

Once you have attached the slim tube HE with dome to the hearing  
instrument, you are ready to place the hearing instrument on the patient’s ear�

1

  

Place the hearing instrument over the top of the ear�

2

 

  Hold the slim tube HE where it attaches to the dome and gently push 

the dome into the patient’s ear canal� The dome should be placed 
far enough into the ear so that the slim tube HE lies flush against the 
head� When looking at your patient face-to-face, you should not see 
the slim tube HE sticking out�

3

   Place the retention piece of the slim tube HE into the concha bowl�

Cleaning the slim tube HE and dome.

The slim tube HE and dome need to be changed every three months or when 

they become stiff or brittle� As part of the fitting, you will need to counsel the 
patient on how to clean the dome and the inside of the slim tube HE� Patients 
should wipe the dome daily with a damp cloth� They should also clean the slim 
tube HE periodically, with the cleaning pin provided, when they begin to notice 
debris in and around the tube�

To clean the slim tube HE and dome:

1

   

Follow the instructions for removing the slim tube HE from the  

hearing instrument� See the section »Attaching and removing  
the slim tube HE�«

2

 

  Use a damp cloth to clean the outside of the slim tube HE  

and the dome�

3

 

  Remove the dome from the slim tube HE�

4

 

  Using the cleaning pin provided in the kit, gently insert the cleaning pin 

where the slim tube HE attaches to the hearing instrument and push the 
pin all the way through the tube�

5

 

  Attach the dome to the slim tube HE and attach the slim tube HE to 

the hearing instrument�

   Note: If the retention piece is too long, you may cut the length slightly. Be careful not to shorten the 

length too much, as it could affect retention in the ear.

   Note: The slim tube HE and dome should never be rinsed or submerged in water as water drops may 

become lodged in the tube, block sound or damage the electrical components of the hearing instrument.

1

2

3

Summary of Contents for Slim Tube HE

Page 1: ...Fitting guide Slim Tube HE for BTE products Slim Tube HE complete guide for BTE products ...

Page 2: ...aring instrument 4 Slim tube HE sizes 4 Measuring for an accurate slim tube HE length 5 6 Choosing the correct domes 7 Attaching and removing domes 7 Attaching and removing the slim tube HE 8 Modifying the slim tube HE to improve fit 8 Positioning the hearing instrument on the ear 9 Cleaning the slim tube HE and dome 9 ...

Page 3: ...nt The guide will introduce you to the parts of the system how to measure for the right tube length choose the correct dome size and how to attach the slim tube HE to the hearing instrument so you are ready to program in fitting software Warning to hearing healthcare professional Domes should never be fitted on patients with perforated eardrums exposed middle ear cavities or surgically altered ear...

Page 4: ...3 Slim Tube HE 4 Dome 5 Retention piece Slim tube HE sizes Slim tube HE is available in four different lengths to meet the needs of your patients The red or blue numeric code on the slim tube identifies the ear to which it belongs blue left ear red right ear as well as the tube length Refer to the following chart to identify the various slim tubes that are available Tube Length Code on Tube Short ...

Page 5: ...olor Code Markings blue left ear and red right ear color code markings to indicate which of the patient s ears to measure Horizontal Line Markings measurements must be made horizontally with the measurement tool touching the skin in the gap between the head and the top of the pinna Four Horizontal Sightlines 0 1 2 3 to indicate which tube length you should use for an accurate fitting Measuring for...

Page 6: ...izes can vary and thus require different slim tube HE lengths To measure for an accurate slim tube HE length 1 Place your head at the same horizontal level as the patient s head so you are looking straight into the patient s ear that you wish to fit with the hearing instrument 2 Hold the measurement tool with the appropriate blue or red color code marking for the left or right ear with your thumb ...

Page 7: ...toscopic examination of the patient s ear canal Based on the exami nation choose an appropriate dome to fit the ear canal Good fit Too small Open Dome Choose for Instant fit Open fit Closed Dome Choose for Instant fit Semi occluded fit Power Dome Choose for Instant fit Occluded fit S M L S M L S M L To remove the dome grasp the dome between the thumb and forefinger and pull it off the slim tube HE...

Page 8: ...a low setting for a few seconds 3 Do not use excessive heat to avoid damage to the tube 4 Bend the tubing into the desired shape and hold until it has cooled off Once completely cooled the tube will maintain its new shape Attaching and removing the slim tube HE To attach the slim tube HE 1 Unscrew the earhook from the hearing instrument 2 Hold the slim tube with the dome attached in one hand and t...

Page 9: ...HE Patients should wipe the dome daily with a damp cloth They should also clean the slim tube HE periodically with the cleaning pin provided when they begin to notice debris in and around the tube To clean the slim tube HE and dome 1 Follow the instructions for removing the slim tube HE from the hearing instrument See the section Attaching and removing the slim tube HE 2 Use a damp cloth to clean ...

Page 10: ...DC HAN Hansaton Akustik GmbH Sachsenkamp 5 20097 Hamburg Telefon 49 40 29 80 11 0 Fax 49 40 29 80 11 860 Email info hansaton de Internet www hansaton com ...

Reviews: