Section 2: Software Installation, System Operation, Data Handling
© 2017 Hansatech Instruments Ltd
The Print graph function prints all of the graphics information which is displayed within the
boundary of the data recording area plus event markers whilst the Print params functions outputs
each of the parameters in a single column in descending order as they were recorded.
The size of the print output can also be adjusted to fit different areas on the page with the Set
printer page feature (Figure 2-27).
Fig. 2-27
. Set printer page dialogue box.
The default settings give minimal margins. The print area may be changed by altering the height
and width settings with a constant height / width aspect ratio maintained by ticking the “Auto
Height” option box. The printer page settings apply to output from both Print graph and Print
params functions.
3.6
Instrument Programming Via User-Defined Scripts
The FMS has been developed in parallel with Hansatech Scripting Language (HSL). The latter
consists of a simple, yet extremely versatile, method for pre-programming a maximum of six user-
defined control event / measurement sequences (referred to as protocols or scripts) into the
instrument. Once programmed the protocols are available for operation by a single-key press
under “Local mode” outside of the laboratory or from the Windows software. The Script function
on the “drop-down” menu allows access to the Script Editor for creation / editing of protocols, the
Script Recorder which logs commands used in “PC mode” to automatically generate a script and
the script Run command for Windows operation.
3.6.1
The Script Recorder
The principles behind the Script Recorder are very similar to those of “Macro Record” functions
found in spreadsheet packages. When active the recorder logs the sequence and timing of every
command that the user issues during the experiment and automatically creates a script which
during play-back precisely replicates the original series and timing of commands that were issued.
Summary of Contents for FMS1
Page 1: ......
Page 7: ...Section 1 Measurement Principles and Hardware ...
Page 21: ...Section 2 Software Installation System Operation Data Handling ...
Page 70: ...70 FMS 1 Fluorescence Monitoring System 2017 Hansatech Instruments Ltd ...
Page 71: ...Section 3 Stand Alone Instrument Operation and Data Upload ...
Page 83: ...83 Section 3 Stand Alone Instrument Operation and Data Upload 2017 Hansatech Instruments Ltd ...