
12
• If the reading is not close to the selected buffer, “
WRONG
”
and “
WRONG
” will blink alternatively;
• If the reading is close to the selected buffer and the reading is
stable, the “
READY
” symbol is displayed and the “
CFM
” symbol
starts blinking on the LCD, asking for confirmation.
• Press the
CFM
key: the value is stored into memory and the meter
returns to normal mode.
Notes:
• The meter automatically skips the buffer used for the first
calibration point to avoid erroneous procedure. A difference
of at least 1.5 pH unit is required between the two buffers
used for the offset and slope calibration: once calibrated at
either pH 7.01 or 6.86, the instrument automatically
ignores the other value for the second point (same for pH
10.01 and 9.18).
• During calibration, the secondary LCD shows the selected
buffer value. For the
HI 9125
model, it is possible to
display the buffer temperature during calibration by
pressing
RANGE
.
• To clear a previous calibration and return to the default
values, in calibration mode press
CFM
, then
CAL
before
the first buffer is accepted. The LCD will show “
CLr CAL
”
for one second, and then will return to normal mode.
ONE-POINT CALIBRATION
For optimum accuracy it is always recommended to perform a
two-point calibration, but for a faster operation a single-point
calibration can be used. pH 7.01 or pH 6.86 (NIST) are normally
used for this purpose, even though the meters can be calibrated with
any of the 5 memorized calibration values.
After calibrating the first point (see above), press the
CAL
key to end
the calibration procedure.