Preliminary Examination
Remove the tester and accessories from the packing
material and examine it carefully. If you require any further
information, please contact Hanna Instruments technical
support team at [email protected].
HI981204
is delivered in a cardboard box and is supplied
with:
•
HI73120
ORP electrode
•
HI700224
ORP test solution, 470 mV, 20 mL sachet (6 pcs.)
•
HI73128
Electrode removal tool
• Storage / Protection cap
• 1.5V batteries (4 pcs.)
• Instrument quality certificate
• Instruction manual
Note:
Save all packing material until you are sure that
the tester works correctly. Any damaged or defective item
must be returned in its original packing material with the
supplied accessories.
General Description & Intended Use
HI981204
is a compact ORP and temperature tester, part
of Hanna Instruments pool-line family. It features a two-
buttons operation system and is easy to use. The compact and
waterproof casing is designed to float if accidentally dropped
in the water. The tester is factory calibrated and measures
temperature in both °C and °F. It displays a stability tag that
will disappear once the reading has stabilized.
Battery level and low-battery indicator are displayed on the
LCD to alert the user in the event that low battery power could
adversely affect the readings.
BEPS (Battery Error Prevention System)
The tester will automatically shut off if there is not enough
power to guarantee an accurate reading.
ORP Probe
HI981204
is supplied together with
HI73120
replaceable
ORP electrode.
Temperature Sensor
The stainless steel temperature sensor facilitates faster and
more accurate temperature measurement.
Specifications
Range
±1000 mV
-5.0 to 60 °C (23.0 to 140.0 °F)
Resolution
1 mV
0.1 °C (0.1 °F)
Accuracy
@ 25 °C (77 °F)
±2 mV
±0.5 °C (±1.0 °F)
Calibration
Factory calibrated
Electrode
HI73120
ORP electrode
Battery type
1.5 V
Battery life
Approximately 300 hours of use
Auto-off
After 8 minutes
Environment
-5 to 50 °C (23 to 122 °F); RH
100%
Dimensions
171 x 41 x 26 mm (6.7 x 1.6 x 1.0“)
Weight
83 g (2.9 oz.)
Functional Description
Liquid Crystal
Display
SET/HOLD
button
Battery
compartment
⏻
/MODE
button
Temperature
sensor
HI73120
ORP electrode
LCD Display
1. Stability tag
2. Battery life percentage indicator
3. Low battery indicator
4. Measurement unit
5. First LCD line
6. Second LCD line
Operational Guide
Turning the Tester ON / OFF
Press and hold
⏻
/MODE. All LCD segments will be displayed
for a few seconds followed by battery life percentage indicator
(e.g. % 100 BATT).
To turn the tester OFF, from measurement mode, press
⏻
/
MODE.
“OFF” will be displayed on the second LCD line and the tester
will turn off.
HOLD Mode
From measurement mode, press and hold SET/HOLD until
HOLD is displayed. mV reading will be frozen on the LCD.
Press either button to resume measurement mode.
Setup
Setup mode allows the selection of temperature unit.
To enter the Setup mode, press
⏻
/MODE until “TEMP“ and
the current temperature unit are displayed on the second LCD
line (e.g. “TEMP °C”).
Use SET/HOLD to change the temperature unit, and press
⏻
/
MODE to return to measurement mode.
Measurement
Place the electrode in the solution to be tested while stirring
it gently. The measurement should be taken after the stability
tag ( ) has disappeared.
The ORP (mV) value is displayed on the first LCD line and the
temperature on the second LCD line.
Note:
If measurements are taken successively, rinse
the probe thoroughly in distilled or deionized water to
eliminate cross-contamination. After cleaning, rinse the
probe with some of the sample to be measured.