Maintenance

Please read the information below to ensure the highest 
possible accuracy.
• If measurements are taken successively, rinse the probe 

thoroughly in distilled or deionized water to eliminate 
cross-contamination.

• Clean the electrode monthly using ORP cleaning solution. 

Keep the electrode in 

HI70614

 General purpose cleaning 

solution for 30 minutes and rinse the probe thoroughly with 
water.

• When not in use, add a few drops of 

HI703004

 Storage 

solution to the protective cap. Never store the probe in distilled 
or deionized water.

Electrode Replacement

Use supplied removal tool (

HI73128

) to replace the ORP 

electrode.
1.  Insert the tool into the probe cavity.

2.  Rotate the tool counterclockwise.

3.  Pull the electrode out by using the other side of the tool.

4.  Insert a new ORP electrode following the instruction steps 

in reverse order.

Battery Replacement

Battery life percentage indicator is displayed at power on. If 
the battery level drops below 5%, the   symbol lights up 
indicating that the batteries should be replaced soon.
If the battery level is not adequate to guarantee an accurate 
reading, the tester automatically turns off. 

To replace the batteries, follow the next steps:
1.  Turn OFF the tester.
2.  Remove the four screws on the top of the tester to open the 

battery compartment (figure 1).

3.  Remove the old batteries.
4.  Insert four new 1.5V batteries in the battery compartment 

while paying attention to the correct polarity (figure 2).

5.  Close the battery compartment using the four screws.

figure 1

figure 2

Note: 

Only use the battery type specified in the manual. 

Old batteries should be disposed in accordance with local 

regulations.

Accessories

Electrodes

HI73120

Replaceable ORP electrode

HI73128

Electrode removal tool

ORP Test Solution

HI7021M

240 mV @25 °C, ORP test solution, 230 mL

HI70224L

470 mV @25 °C, ORP test solution, 500 mL

Electrode Cleaning Solution

HI70614L

General purpose cleaning solution, 500 mL

Electrode Storage Solution

HI703004L

Electrode storage solution, 500 mL

Pretreatment Solution

HI7091L

Pretreatment reducing solution

HI7092M

Pretreatment oxidizing solution, 230 mL

Other Accessories

HI740026P

Replacement 1.5 V batteries (12 pcs.)

Certification

All Hanna Instruments conform to the 

CE European Directives.

 

 

Disposal of Electrical & Electronic Equipment.

 The product 

should not be treated as household waste. Instead hand it over 

to the appropriate collection point for the recycling of electrical 

and electronic equipment which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do 

not dispose of them with other household waste. Hand them over 

to the appropriate collection point for recycling.

Ensuring proper product and battery disposal prevents potential 

negative consequences for the environment and human health. 

For more information, contact your city, your local household 

waste disposal service, the place of purchase or go to www.

hannainst.com.

Recommendations for Users

Before using this tester, make sure that it is entirely suitable for 

your specific application and for the environment in which it is 

used. Avoid touching the electrode at all times. Any variation 

introduced by the user to the supplied equipment may degrade 

the tester’s performance. For your and the tester’s safety do not 

use or store the tester in hazardous environments.

Warranty

HI981204

 is warranted for a period of one year against defects 

in workmanship and materials when used for its intended 

purpose and maintained according to instructions. The electrode 

is warranted for a period of six months. This warranty is 

limited to repair or replacement free of charge. Damage due to 

accidents, misuse, tampering or lack of prescribed maintenance 

is not covered. If service is required, contact your local Hanna 

Instruments Office. If under warranty, report the model number, 

date of purchase, serial number and the nature of the problem. 

If the repair is not covered by the warranty, you will be notified 

of the charges incurred. 

If the instrument is to be returned to Hanna Instruments Office, 

first obtain a Returned Goods Authorization (RGA) number from 

the Technical Service department and then send it with shipping 

costs prepaid. When shipping any instrument, make sure it is 

properly packaged for complete protection.

Hanna Instruments reserves the right to modify the design, 

construction, or appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited 

without the written consent of the copyright owner, Hanna Instruments 

Inc., Woonsocket, Rhode Island, 02895, USA.

IST981204 12/20

HI981204

Waterproof ORP 

& Temperature Tester

INSTRUCTION MANUAL

Summary of Contents for Pool Line HI981204

Page 1: ...HI981204 Waterproof ORP Temperature Tester INSTRUCTION MANUAL ...

Page 2: ...6 pcs HI73128 Electrode removal tool Storage Protection cap 1 5V batteries 4 pcs Instrument quality certificate Instruction manual Note Save all packing material until you are sure that the tester works correctly Any damaged or defective item must be returned in its original packing material with the supplied accessories General Description Intended Use HI981204 is a compact ORP and temperature te...

Page 3: ... temperature sensor facilitates faster and more accurate temperature measurement Specifications Range 1000 mV 5 0 to 60 C 23 0 to 140 0 F Resolution 1 mV 0 1 C 0 1 F Accuracy 25 C 77 F 2 mV 0 5 C 1 0 F Calibration Factory calibrated Electrode HI73120 ORP electrode Battery type 1 5 V Battery life Approximately 300 hours of use Auto off After 8 minutes Environment 5 to 50 C 23 to 122 F RH 100 Dimens...

Page 4: ... Display SET HOLD button Battery compartment MODE button Temperature sensor HI73120 ORP electrode LCD Display 1 Stability tag 2 Battery life percentage indicator 3 Low battery indicator 4 Measurement unit 5 First LCD line 6 Second LCD line ...

Page 5: ...lection of temperature unit To enter the Setup mode press MODE until TEMP and the current temperature unit are displayed on the second LCD line e g TEMP C Use SET HOLD to change the temperature unit and press MODE to return to measurement mode Measurement Place the electrode in the solution to be tested while stirring it gently The measurement should be taken after the stability tag has disappeare...

Page 6: ... protective cap Never store the probe in distilled or deionized water Electrode Replacement Use supplied removal tool HI73128 to replace the ORP electrode 1 Insert the tool into the probe cavity 2 Rotate the tool counterclockwise 3 Pull the electrode out by using the other side of the tool 4 Insert a new ORP electrode following the instruction steps in reverse order Battery Replacement Battery lif...

Page 7: ...pecified in the manual Old batteries should be disposed in accordance with local regulations Accessories Electrodes HI73120 Replaceable ORP electrode HI73128 Electrode removal tool ORP Test Solution HI7021M 240 mV 25 C ORP test solution 230 mL HI70224L 470 mV 25 C ORP test solution 500 mL Electrode Cleaning Solution HI70614L General purpose cleaning solution 500 mL Electrode Storage Solution HI703...

Page 8: ...se or store the tester in hazardous environments Warranty HI981204 is warranted for a period of one year against defects in workmanship and materials when used for its intended purpose and maintained according to instructions The electrode is warranted for a period of six months This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of pre...

Reviews: