SPECIFICATIONS:

RANGE

0.0 to 14.0 pH

RESOLUTION

0.1 pH

ACCURACY 

(@20°C/68°F)

±0.1 pH

TYPICAL EMC DEVIATION

±0.1 pH

ENVIRONMENT

0 to 50 C

(32 to 122 F); 95% RH

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

Manual, at two points through

offset and slope trimmers

BATTERY TYPE

 4 x 1.5V alkaline (included)

BATTERY LIFE

approx. 1700 hours

of continuous use

DIMENSIONS

175 x 41 x 23 mm

(7.9 x 1.8 x 1")

WEIGHT

78 g (2.7 oz.)

ACCESSORIES:

HI 73106

Spare electrode

HI 77400P

pH 4.01 & 7.01 calibration kit
(5 x 20 mL, each)

HI 77700P

pH 7.01 calibration kit
(10 x 20 mL)

HI 770710P

pH 7.01 & 10.01 calibration kit
(5 x 20 mL, each)

HI 7004M

pH 4.01 buffer solution
(230 mL bottle)

HI 7007M

pH 7.01 buffer solution
(230 mL bottle)

HI 7010M

pH 10.01 buffer solution
(230 mL bottle)

HI 70300M

Storage solution (230mL)

HI 7061M

Electrode cleaning solution
(230mL bottle)

HI 731326

Calibration screwdriver (20 pcs)

IS98107-R3       09/01

ELECTRODE REPLACEMENT:

The electrode can be easily replaced in the following
way:
• Slide off the battery cover.
• Remove the screw on the back of the pHep

pHep

pHep

pHep

pHep

located below the battery compartment and slide
off the whole internal part of the tester.

• To remove the electrode proceed as follows:

Remove the fastening screw on the front (1) to
loose the electrode. The internal circuit is con-
nected to the electrode through two wires (one
for the glass sensor and one for the reference).
Remove the two small screws (2 & 3) which
fasten the two connecting wires to their sock-
ets.

• Remove the electrode and

replace it with a new
HI73106.

• Connect the new electrode

as follows:
first fasten the screw on the
front (1) to attach the elec-
trode to the circuit. Insert
the two small wires into
their sockets and fasten the
two screws (2 & 3).

• The new electrode is supplied with four new O-

rings (4). Make sure they are placed properly
(as shown on the above drawing) before
reinserting the circuit into the plastic casing.

• Reinsert the internal circuit with the new

electrode into the plastic casing.

• Fasten the screw on the back below the

battery compartment.

Internal circuit and electrode:

• Reinsert the battery compartment cover.

• Perform a new calibration before using the

pHep 

pHep 

pHep 

pHep 

pHep  again.

pHep

®

®

®

®

®

HI 98107

Pocket-sized pH Meter

HI73106

ONE YEAR WARRANTY & SERVICEABLE:

ONE YEAR WARRANTY & SERVICEABLE:

ONE YEAR WARRANTY & SERVICEABLE:

ONE YEAR WARRANTY & SERVICEABLE:

ONE YEAR WARRANTY & SERVICEABLE:
Covered by one year

one year

one year

one year

one year warranty against defects in work-

manship and materials, this tester is now completely
serviceable. Contact your dealer for further information.