Hanna Instruments pH 211 Instruction Manual Download Page 5

8

9

POWER CONNECTION

POWER CONNECTION

POWER CONNECTION

POWER CONNECTION

POWER CONNECTION

Plug the 12 VDC adapter into the power supply socket.

Notes:

• These instruments use non volatile memory to retain the

pH, mV, temperature calibrations and all other settings,
even when unplugged.

• Make sure a fuse protects the main line.

ELECTRODE AND PROBE CONNECTIONS

ELECTRODE AND PROBE CONNECTIONS

ELECTRODE AND PROBE CONNECTIONS

ELECTRODE AND PROBE CONNECTIONS

ELECTRODE AND PROBE CONNECTIONS

For pH or ORP measurements connect an electrode with internal
reference to the BNC connector on the back of the instrument.
For electrodes with a separate reference connect the electrode’s BNC to the
BNC connector and the reference electrode plug to the reference socket.
For temperature measurements and automatic temperature compen-
sation connect the temperature probe to the appropriate socket.

INSTRUMENT START-UP

INSTRUMENT START-UP

INSTRUMENT START-UP

INSTRUMENT START-UP

INSTRUMENT START-UP

• Turn the instrument on by pressing the 

ON/OFF

switch.

• All LCD tags are displayed and a beep is heard

while the instruments perform a self test (

pH 212

and 

pH 213

 only).

pH 210

 & 

pH 211

pH 212

 & 

pH 213

pH  MEASUREMENTS

pH  MEASUREMENTS

pH  MEASUREMENTS

pH  MEASUREMENTS

pH  MEASUREMENTS

Make sure the instrument has been calibrated before

taking pH measurements.

• Submerge the electrode tip and the temperature

probe approximately 4 cm (1½”) into the sample

to be tested and stir gently. Allow time for the

electrode to stabilize.

C

• The pH is displayed on the primary LCD and the temperature on

the secondary LCD.

• If the pH reading is out of range, “----” will be displayed on the

LCD.

If measurements are taken successively in different samples, it is
recommended to rinse the electrode thoroughly with deionized water
or tap water and then with some of the next sample in order to
prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the
pH accurately, the temperature effect must be compensated for.
To use the 

Automatic Temperature Compensation

 feature, connect

and submerge the 

HI 7669/2W

 temperature probe into the sample

as close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known,simple 

manual temperature

compensation

 can be performed by disconnecting the temperature probe.

The display will then show the default temperature
of 25 °C or the last recorded temperature reading
with the “°C” tag blinking. The temperature can now
be adjusted with the 

ARROW

 keys (from -9.9 °C to

120.0 °C).

ORP MEASUREMENTS (

ORP MEASUREMENTS (

ORP MEASUREMENTS (

ORP MEASUREMENTS (

ORP MEASUREMENTS (pppppH 211

H 211

H 211

H 211

H 211 and p

 and p

 and p

 and p

 and pH 213 

H 213 

H 213 

H 213 

H 213 only

only

only

only

only)))))

An optional ORP electrode must be used to perform ORP measurements
(see Accessories).
Oxidation-reduction potential (REDOX) measurements provide the
quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a redox measurement, the surface
of the ORP electrode must be clean and smooth.

• Press 

RANGE

 to enter mV range.

• Submerge the tip of the ORP electrode (4 cm/

1½") into the sample to be tested and allow a
few seconds for the reading to stabilize.

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

C

C

pH

C

Summary of Contents for pH 211

Page 1: ...ffer nd Confirm 1 Buffer st CFM Select 1 Buffer st Start Calibration CAL COMM RS232 Communication MR Memory Recall Select pH ORP RANGE MEM Memorize Reading MR MEM COMM C C RANGE CFM CAL Instruction Ma...

Page 2: ...your Dealer or the nearest Hanna Customer Service Center Each instrument is supplied with HI 1131B Glass body Combination pH Electrode HI 7669 2W Temperature Probe HI 76404 Electrode Holder pH 4 01 7...

Page 3: ...1 2 H p V m 1 C 1 0 Y C A R U C C A F 8 6 C 0 2 H p 1 0 0 V m 2 0 V m 9 9 9 3 V m 1 V m 0 0 0 2 C 0 0 0 1 0 0 C 5 0 1 e d i s t u o C r o r r e e b o r p g n i d u l c x e n o i t a r b i l a C H p s...

Page 4: ...ffer nd Confirm 1 Buffer st CFM Select 1 Buffer st Start Calibration CAL COMM RS232 Communication MR Memory Recall Select pH ORP RANGE MEM Memorize Reading MR MEM COMM C C RANGE CFM CAL E G N A R H p...

Page 5: ...on the secondary LCD If the pH reading is out of range will be displayed on the LCD If measurements are taken successively in different samples it is recommended to rinse the electrode thoroughly wit...

Page 6: ...ccuracy is required The instrument should be recalibrated Whenever the pH electrode is replaced At least once a week After testing aggressive chemicals PREPARATION PREPARATION PREPARATION PREPARATION...

Page 7: ...rn to measurement mode Notes If the value measured by the meter is not close to the selected buffer WRONG and WRONG tags will blink alternately In this case check if the correct buffer has been used o...

Page 8: ...er With the instrument off press and hold down the CAL MEM keys then power on the instrument The CAL tag will appear and the secondary LCD will show 0 0 C Immerse the temperature probe in the vessel w...

Page 9: ...LIBRATION V CALIBRATION V CALIBRATION V CALIBRATION V CALIBRATION p p p p pH 213 H 213 H 213 H 213 H 213 only only only only only for technical personnel only C C CFM COMM All the instruments are fact...

Page 10: ...set to pH Err 6 is sent TM Causes the instrument to send the temperature reading Err 3 is sent if out of range Notes Either small or capital letters can be used The characters sent by the meter are a...

Page 11: ...Storage Solution or in its absence Filling Solution HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 for double junction electrodes Follow the Preparation Procedure on page 20 before taki...

Page 12: ...r r o f y l n o e h t n i e g n a r f o t u O e l a c s V m y r D n o i t c n u j e n a r b m e m r o 0 0 3 0 7 I H n i k a o S e g a r o t s 0 0 3 0 8 I H n o i t u l o s t o n s e o d r e t e m e h...

Page 13: ...FDA approved bottle 4x50 mL for single junction electrodes HI 8072 1M KNO3 Electrolyte in FDA approved bottle 4x50 mL HI 8082 3 5M KCl Electrolyte in FDA approved bottle 4x50 mL Typical Electrode Lif...

Page 14: ...DIA 0 29 S VERSION HI 1331 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing Pretreatment Solution 460 mL bottle HI 7092L Oxidizing Pretreatment Solution 460 mL bottle pH ELECTRODES All electrodes part num...

Page 15: ...12 mm 0 5 S VERSION HI 3131 120 mm 4 7 12 mm 0 5 HI 3230 S VERSION 120 mm 4 7 12 mm 0 5 HI 4430 S VERSION FC 100B Plastic body Kynar double junction refillable combination pH electrode Use general pu...

Page 16: ...E ELECTRODES CONNECT TO THE BNC SOCKET OF THE METER HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3 0 mm 0 12 CABLE WITH BNC Windows is registered Trademark of Microsoft Co RECOMMENDATIONS FOR USERS B...

Page 17: ...02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739...

Reviews: