10
SPECIFICATIONS
PCA 300
PCA 301
Range
0.00 to 5.00 mg/L
0.00 to 5.00 mg/L
Free Cl
2
Total Residual Cl
2
Resolution
0.01 mg/L
Accuracy
±8% of reading or ±0.05 mg/L
whichever is greater
Typical EMC
±0.05 mg/L
Deviation
Minimum
0.05 mg/L
detectable level
Repeatability
±0.05 mg/L
Response time
Related to the sampling time
selected.
Typical for a full scale step change and
5 minutes between two consecutive
samples: one sampling cycle for 90%
response and two sampling cycles for
100% response
Sampling rate
Adjustable from 3 to 102 minutes
Sample inlet
0.07 bar (1psig) min., 4 bar (57.2 psig)
pressure
max. An internal regulator reduces
pressure from 4 bar (57.2 psig) to
1 bar (14.3 psig)
Sample flow
Flow rate of 300 mL/min is recom-
range
mended. Minimum and maximum
allowed are 100 mL/min and
500 mL/min, respectively
Sample
5 to 40°C (41 to 104°F)
temperature
range
Interferences
Oxidizing agents such as: Iodine,
Bromine, Ozone, Chlorine Dioxide,
Permanganate, Hexavalent Chromium.
Hardness must not exceed 1000 mg/L
as CaCO
3
. Alkalinity must not exceed