background image

HI99162

Waterproof  

pH & Temperature Meter  

for Milk

INSTRUCTION MANUAL

Summary of Contents for HI99162

Page 1: ...HI99162 Waterproof pH Temperature Meter for Milk INSTRUCTION MANUAL ...

Page 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this meter This manual will provide you with the necessary information for correct use of this meter as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our worldwide contact list at www hannainst com Dear Customer...

Page 3: ...eral description and Intended use Main features Specifications Display description Operational guide Meter setup Battery replacement Accessories Electrode maintenance Certification Recommendations for users Warranty 4 5 7 9 15 6 8 16 14 17 18 19 19 ...

Page 4: ...s supplied with FC1013 pH temperature probe with DIN connector and 1m 3 3 cable HI70004 pH 4 01 buffer 1 sachet HI70007 pH 7 01 buffer 1 sachet HI700640 electrode cleaning solution for milk deposits 2 sachets 100 mL beaker 1 pcs 1 5V AAA alkaline batteries Instrument quality certificate Electrode quality certificate Instruction manual Note Save all packing material until you are sure that the inst...

Page 5: ...tic pH calibration at one or two points The FC1013 is a specially designed pH electrode intended for the direct determination of pH in milk The FC1013 pH probe has a durable PVDF body that can be safely taken on the dairy farm or production floor It features an open junction reference system that is resistant to clogging from milk based solids The FC1013 has a built in temperature sensor for tempe...

Page 6: ...within two memorized buffer sets standard or NIST Selectable temperature unit C or F Electrode condition indicator mV of pH measurement for electrode check FC1013 dedicated pH with integrated temperature sensor Probe quick connect system Battery life indication and low battery detection Keystroke confirmation tone Auto off function Waterproof casing IP67 ...

Page 7: ...lectable buffer set Standard 4 01 7 01 10 01 or NIST 4 01 6 86 9 18 Probe included FC1013 preamplified pH and temperature probe DIN connector and 1 m 3 3 cable Battery type life 1 5V AAA 3 pcs approx 1400 hours of continuous use Auto off user selectable after 8 min 60 min or disabled Environment 0 to 50 C 32 to 122 F RH max 100 Meter dimensions 154 x 63 x 30 mm 6 1 x 2 5 x 1 2 Weight with batterie...

Page 8: ...emperature Compensation indicator 2 Stability indicator 3 Battery percentage 4 Electrode condition indicator 5 Low battery indicator 6 Measurement unit 7 Primary LCD 8 Temperature unit 9 Secondary LCD 10 Meter modes indicator 11 pH calibration buffer s used ...

Page 9: ...led in their place The meter is provided with an active acoustic signal when a key is pressed At start up the meter displays all LCD segments for a few seconds followed by the percentage indicationoftheremainingbatterylife displaying WAIT until electrode check is in process then the meter enters the normal measurement mode Battery life Measurement mode LCD check up Electrode check Note The meter d...

Page 10: ...ELECTING MEASUREMENT RANGE While in measurement mode press the SET button to select pH or pH mV measurement on the first LCD line FREEZING MEASUREMENT VALUES While in measurement mode press and hold the SET button until HOLD appears on the secondary LCD The HOLD remains for 1 second and reading of pH mV and temperature will be frozen on the LCD with H blinking Press any button to resume active mea...

Page 11: ... the first calibration buffer If performing a two point calibration use pH 7 01 pH 6 86 for NIST buffer first The meter will enter the calibration mode displaying pH 7 01 USE and tag blinking or pH 6 86 USE for NIST Follow directions for single and two point calibration below Single point calibration 1 Place the probe in any buffer from the selected buffer set The meter will automatically recogniz...

Page 12: ...t calibration For better accuracy it is always recommended to carry out a two point calibration After the calibration procedure has been completed the tag is turned on together with the calibrated points Exiting calibration and resetting default values After entering the calibration mode and before the first point is accepted it is possible to quit the procedure and return to the last calibration ...

Page 13: ...ere is still only 1 bar or 1 bar blinking replace the probe Sensor check Setting the meter to pH mV range the user can check the sensor status at any time The offset value is the reading in pH 7 01 buffer 25 C 77 F If this reading is outside the range 30 mV the electrode is considered very poor The slope value of the sensor is the difference between readings in pH 7 01 and in pH 4 01 buffers When ...

Page 14: ...Press to confirm and to enter the BEEP selection To switch the beep tone ON or OFF press the SET button press to confirm and to enter the calibration buffer selection pH 7 01 BUFF To change the buffer set the meter will show the current buffer set pH 7 01 BUFF for standard buffer set 4 01 7 01 10 01 or pH 6 86 BUFF for NIST buffer set 4 01 6 86 9 18 Change the set with the SET button Press to conf...

Page 15: ... display shows bAtt DEAD for a few seconds then the meter powers off Immediately replace the batteries with new ones The batteries are accessed by opening the battery cover on the back of the instrument Remove protective boot if present Replace the three 1 5V AAA alkaline batteries located in the battery compartment observing the indicated polarity Replace the battery cover making sure that the ga...

Page 16: ...H electrode storage solution 500 mL HI7061L pH electrode cleaning solution 500 mL HI70640L Cleaning solution for milk deposits 500 mL HI7082L pH electrode fill solution 500 mL HI710029 Blue silicon rubber boot HI710142 Black carrying case for HI991XX portable instruments HI76405 Electrode holder HI77400P Calibration kit pH 4 and pH 7 20 mL 5 pcs each ACCESSORIES Code and serial number on the other...

Page 17: ...ould be kept moist and not allowed to dry Replace protective cap with a few drops of HI70300 Storage Solution Follow PREPARATION above before taking measurements Replace the fill hole cover Note Never store the electrode in distilled water PERIODIC MAINTENANCE Inspect the electrode for any scratches or cracks If any present replace the electrode Rinse off any salt deposits with water Follow the ST...

Page 18: ...onic equipment which will conserve natural resources Disposal of waste batteries This product contains batteries do not dispose of them with other household waste Hand them over to the appropriate collection point for recycling Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health For more information contact your city your local...

Page 19: ...ed by the warranty you will be notified of the charges incurred If the meter is to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid When shipping any meter make sure it is properly packed for complete protection Warranty Before using this product make sure it is entirely suitable f...

Page 20: ...Printed in ROMANIA World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN99162 ...