OPERATION

• Remove the protective cap and turn 

ORP

 on with by sliding the ON/OFF 

switch on the top of the meter.

• Immerse it into the solution to

be tested, without exceeding the
maximum immersion level.

• Stir gently and wait for stable reading.

• After use, rinse the electrode with tap water and

replace the protective cap.

MAINTENANCE

• Check the meter by im-

mersing it in 

HI7021

 test

solution. The reading should
be between 240±20 mV at 
25

 ºC (

77

 ºF).

Otherwise clean the electrode 
platinum tip by rubbing it
with a lint-free cloth soaked
with 

HI7061

 cleaning solu-

tion.

• For more accurate readings, condition the meter 

by immersing it in a pretreatment reducing

(

HI7091

)  or  oxidizing  (

HI7092

)  solution  for 

about 30 minutes before taking measurements.

• In case of erroneous readings even after an

accurate conditioning and calibra-
tion, the reference junction might
be contaminated or clogged.

•  Pull  out  2  mm  (1/8’’) 

of the cloth junction
to renew the electrode

reference  (it  is  recom

-

mended to cut the cloth
leaving always at least

2  mm  -1/8’’  over  the 
reference compartment).

  Repeat the maintenance and conditioning 

procedure. 
The cloth junction can be pulled out approxi-
mately 20 times. After that, the electrode should 
be replaced.

ELECTRODE REPLACEMENT

For  replacing  the  electrode,  contact  your  local 

Hanna Instruments Office.

BEPS & BATTERY REPLACEMENT

The  meter  is  provided  with  BEPS  (Battery  Error 
Prevention  System)  which  avoid  any  erroneous 

reading due to low battery level, by automatically 
switching the meter off.
To replace the batteries, slide off the battery 
compartment cover and replace all four 1.5V 
alkaline batteries, while paying attention to the 
correct polarity. 

Batteries should only be replaced in a non-
hazardous area and using the battery type specified 
in this instruction manual. 