METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description
Quantity
HI95762-0
Free Chlorine ULR Reagent
1 packet
REAGENT SETS
HI95762-01
Free Chlorine ULR Reagent - 100 tests
HI95762-03
Free Chlorine ULR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample
(up to the mark). Replace the plastic stopper and
the cap.
• Insert the cuvette into the holder and ensure that
the notch on the cap is positioned securely in the
groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add the content of one packet of
HI95762-0
Free
Chlorine ULR Reagent. Replace the plastic stopper
and the cap. Shake gently for about 20 seconds.
10 mL