
21
METHOD PROCEDURE
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Read
and the display will show a 45 second countdown prior to the
measurement. To skip the timer, press
Read
twice. When the timer ends, the
meter will perform the reading. The instrument displays the results in
mg/L
of
copper (Cu).
INTERFERENCES
Interference may be caused by:
• Cyanide, Silver
• For samples overcoming buffering capacity of reagent around pH 6.8, pH should be
adjusted between 6 and 8