METHOD PROCEDURE
• Press
Zero
and the display will display a 15 minutes countdown. To skip the timer
press
Zero
twice.
• Add one packet of
HI93740C-0
Nickel LR
Reagent C to each beaker, cap and swirl to mix
until completely dissolved.
• Fill one cuvette (#1) with 10 mL of the blank
(up to the mark). Replace the plastic stopper
and the cap.
• Insert the cuvette into the holder and ensure
that the notch on the cap is positioned securely
in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Fill a second cuvette (#2) with 10 mL of the
reacted sample (up to the mark). Replace the
plastic stopper and the cap.
#1
#2
#1
10 mL
#1
#2
10 mL