23
METHOD PROCEDURE
• Insert the blank (#1) into the holder and ensure
that the notch on the cap is positioned securely in
the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Fill second cuvette (#2) with 10 mL of the reacted
sample (up to the mark). Replace the plastic
stopper and the cap.
• Insert the cuvette into the holder and ensure that
the notch on the cap is positioned securely in the
groove.
• Press
Read
to start the reading. The instrument displays the results in
mg/L
of
chlorine dioxide (ClO
2
).
#1
#2
10 mL
#2