METHOD PROCEDURE
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add 0.5 mL of
HI93731B-0
Zinc Reagent B.
Replace the plastic stopper and the cap. Swirl
gently for 15 seconds.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Read
. The display will show a 3 minute and 30 second countdown prior to the
measurement. To skip the timer, press
Read
twice. When the timer ends, the meter
will perform the reading. The instrument displays the results in
mg/L
of
zinc (Zn).
INTERFERENCES
Interference may be caused by:
• Iron above 7 mg/L
• Aluminum above 6 mg/L
• Copper, Manganese, Nickel above 5 mg/L
• Cadmium above 0.5 mg/L