19
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description
Quantity
HI93728-0
Nitrate Reagent
1 packet
REAGENT SETS
HI93728-01
Nitrate Reagent - 100 tests
HI93728-03
Nitrate Reagent - 300 tests
For other accessories see page 24.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of sample (up to the
mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press the
Zero
key. The display will show
“-0.0-”
when the meter is zeroed and
ready for measurement.
• Remove the cuvette and add the content of one packet
of
HI93728-0
Nitrate Reagent. Replace the plastic
stopper and the cap.
10 mL