21
METHOD PROCEDURE
• Add one packet of
HI93722-0
Cyanuric Acid Reagent
and mix to dissolve.
• Fill a second cuvette with 10 mL of the reacted sample
(up to the mark). Replace the plastic stopper and the
cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Read
and the display will show a 45 second countdown prior to the
measurement. To skip the timer, press
Read
twice. When the timer ends the meter
will perform the reading. The instrument displays the results in
mg/L
of
cyanuric
acid.
10 mL
# 2