20
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description
Quantity
HI93717A-0
Phosphate HR Reagent A
10 drops
HI93717B-0
Phosphate HR Reagent B
1 packet
REAGENT SETS
HI93717-01
Phosphate HR Reagent - 100 tests
HI93717-03
Phosphate HR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add 10 drops of
HI93717A-0
Phosphate HR
Reagent A.
10 mL
×
10