METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description Quantity
HI93708-0
Nitrite HR Reagent
1 packet
REAGENT SETS
HI93708-01
Nitrite HR Reagent - 100 tests
HI93708-03
Nitrite HR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add one packet of
HI93708-0
Nitrite High
Range Reagent. Replace the plastic stopper
and the cap. Shake gently until completely
dissolved.
10 mL