20
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description Quantity
HI93707-0
Nitrite LR Reagent
1 packet
REAGENT SETS
HI93707-01
Nitrite LR Reagent - 100 tests
HI93707-03
Nitrite LR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add the content of one packet of
HI93707-
0
Nitrite LR Reagent. Replace the plastic
stopper and the cap. Shake gently for about
15 seconds.
10 mL