HI96752

 is warranted for two years against defects in workmanship and 

materials when used for its intended purpose and maintained according 
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed main-
tenance are not covered. 
If service is required, contact your local Hanna Instruments Office. If under 
warranty, report the model number, date of purchase, serial number and 
the nature of the failure. If the repair is not covered by the warranty, you 
will be notified of the charges incurred. 
If the instrument is to be returned to Hanna Instruments, first obtain a 
Returned Goods Authorization Number from the Customer Service Depart-
ment and then send it with shipment costs prepaid. When shipping any 
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card 
within 14 days from the date of purchase.

WARRANTY

IST96752   01/18

Hanna Instruments reserves the right to modify the design, construction, or  

appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the 

written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, 

Rhode Island, 02895, USA

Reagent Sets

HI93752-01

Reagents for 100 tests (50 tests Ca and 50 tests Mg)

HI93752-03

Reagents for 300 tests (150 tests Ca and 150 

tests Mg)

Other Accessories

HI96752-11

CAL Check™

 Standard Cuvettes for Calcium  

(1 set)

HI96754-11

CAL Check™

 Standard Cuvettes for Magnesium 

(1 set)

HI740029P

9V battery (10 pcs.)

HI731318

Tissue for wiping cuvettes (4 pcs.)

HI731331

Glass cuvettes (4 pcs.) 

HI731335

Caps for cuvettes (4 pcs.)

HI93703-50

Cuvettes cleaning solution (230 mL)

ACCESSORIES

Please examine this product carefully. Make sure that the instrument is not 

damaged. If any damage occured during shipment, please contact your local 

Hanna Instruments Office.
Each 

HI96752

 Ion Selective Meter is supplied complete with: 

• Sample Cuvettes and Caps (2 pcs.)
• 9V Battery
• Instruction Manual
• Quality Certificate

Note: Save all packing material until you are sure that the instrument 

works correctly. Any defective item must be returned in its original 

packing.

1.  

RANGE/GLP

 

key: press to change the paramter, press and hold 

for three seconds to enter GLP mode. In calibration mode press to edit 

the date and time.

2. 

CAL CHECK

 key: press to perform the validation of the meter, or press 

and hold for three seconds to enter calibration mode. 

3. 

ZERO/CFM

 key: press to zero the meter prior to measurement, to 

confirm edited values or to confirm factory calibration restore.

4. 

READ /TIMER

 key: In measurement mode, press to make a 

measurement, or press and hold for three seconds to start a pre-

programmed countdown prior to measurement. In GLP mode press to 

view the next screen.

5. 

ON/OFF

 key: to turn the meter on and off.

6.  Liquid Crystal Display (LCD)

7.  Cuvette alignment indicator

8.  Cuvette holder

 1. The measuring scheme (lamp, cuvette, detector), appears during 

different phases of zero or reading measurement

 2. Error messages and warnings

 3. The battery icon indicates the charge state of the battery

 4. The hourglass appears when an internal check is in progress

 5. Status messages

 6. The chronometer appears when the reaction timer is running

 7. The month, day and date icons appear when a date is displayed

 8. Four digit main display

 9. Measuring units

10.  Four digit secondary display

Thank You

Thank you for choosing a Hanna Instruments product. Please read 
this instruction manual carefully before using the instrument. 
For more information about Hanna Instruments and our products, 
visit www.hannainst.com.
For technical support, contact your local Hanna Instruments 
Office or e-mail us at [email protected]
Find your local Hanna Instruments Office at www.hannainst.com

PRELIMINARY EXAMINATION

 

For more details about spare parts and accessories see 

“Accessories”.

SPECIFICATIONS

Range

Calcium       0 to 400 mg/L   
Magnesium  0 to 150 mg/L     

Resolution

1 mg/L

Accuracy 
@25 

°

C (77 

°

F)

Calcium       ±10 mg/L ±5% of reading
Magnesium  ±3 mg/L ±3% of reading

Light source

Light Emitting Diode

Light Detector

Silicon Photocell with narrow band interference 
filter @466 nm

Method

For Calcium: Adaptation of the Oxalate method. 
For Magnesium: Adaptation of the Calmagite 
method.

Environment

0 to 50 

°

C (32 to 122 

°

F); 

max 95% RH non-condensing

Battery Type

9V (1 pc.)

Auto-Shut off

After 10’ of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)

Weight

320 g (11.3 oz.)

FUNCTIONAL DESCRIPTION

DISPLAY ELEMENTS DESCRIPTION:

Dead battery:

 This indicates that the battery is 

dead and must be replaced. Once this indication 
is displayed, normal operation of the instrument 
will be interrupted. Change the battery and 
restart the meter.

On sample reading:

During calibration procedure:

Other errors and warnings:

Cooling lamp:

 The instrument waits for the 

lamp to cool down.

Standard High:

 The standard reading is higher 

than expected.

On zero reading:

ERRORS AND WARNINGS

Battery low:

 The battery must be replaced soon.

3•

 

Calcium:

 Using a 5 mL syringe add exactly 

3 mL of sample to the cuvette. Use the pi-

pette to fill the cuvette up to the 10 mL mark 

with the 

HI93752A-Ca

 Buffer Reagent. Add  

4 drops of Buffer Reagent. 

 

Magnesium

: Using a 1 mL syringe add 

exactly 1 mL of 

HI93752A-Mg

 Buffer Re-

agent. Use the pipette to fill the cuvette up 

to the 10 mL mark with the 

HI93752B-Mg 

Indicator Reagent. 

4•

  Replace the cap and invert several times 

to mix. Place the cuvette into the holder 

and ensure that the notch on the cap is 

positioned securely into the groove.

5•

 Press 

ZERO/CFM

 and the lamp, cuvette and 

detector icons will appear on the display, 

depending on the measurement phase.

6•

  After a few seconds the display will show      

“-0.0-”

. The meter is now zeroed and ready 

for measurement.

7•

  Remove the cuvette.

8•

  

Calcium

: Using a 1 mL syringe add 

exactly 1 mL of the 

HI93752B-Ca

 Oxalate 

reagent. 

Magnesium

: Using another 1 mL 

syringe, add to the cuvette exactly 0.5 mL 

of sample. 

Note: Do not mix up the two 1 

mL syringes!

9•

 

Calcium

: Replace the cap and invert the 

cuvette 10 times to mix.

 

Magnesium

: Replace the cap and invert 

several times to mix. 

10• 

Replace the cuvette into the holder and en-

sure that the notch on the cap is positioned 

securely into the groove. 

11• 

Press 

READ /TIMER

 for three seconds. 

The display will show the countdown 

prior to measurement. The beeper is 

playing a beep at the end of countdown 

period. Calcium only: Invert cuvette 10 

times again before the timer expires. 

Alternatively, wait for 5 minutes then 

invert the cuvette 10 times and press  

READ /TIMER.

   

Magnesium only: Alternatively, wait for  

15 seconds and press 

READ /TIMER

. In 

all cases the lamp, cuvette and detector icons 

will appear on the display, depending on the 

measurement phase.

12•

 The instrument directly displays concentration 

in mg/L of Calcium or Magnesium on the 

LCD, depending on the selected parameter.

11

3

2

4

5-6

8-9

HI93752B

Ca

10

MEASUREMENT PROCEDURE

Measurement

 

 or

 or

 or

12

INTERFERENCES

:

• 

Calcium:

 

Concentration

 

Reading

Acidity (as CaCO

3

above 1000 mg/L  lower

Alkalinity (as CaCO

3

)  above 1000 mg/L  higher

Magnesium (Mg

2+

above 400 mg/L 

higher

• 

Magnesium:

 

Concentration

 

Reading

Acidity (as CaCO

3

above 1000 mg/L  lower

Alkalinity (as CaCO

3

)  above 1000 mg/L  higher

Calcium (Ca

2+

above 200 mg/L 

higher

Iron ,  a 

higher

Aluminum - 

higher

Copper -  higher

No Light:

 The instrument cannot adjust the 

light level. Please check that the sample does 
not contain any debris.

Light Low:

 There is not enough light to perform 

a measurement. Please check the preparation of 
the zero cuvette.

Light High:

 There is too much light to perform 

a measurement. Please check the preparation of 
the zero cuvette.

Inverted cuvettes:

 The sample and the zero 

cuvette are inverted.

Zero:

 A zero reading was not taken. Follow the 

instructions of the measurement procedure for 
zeroing the meter. 

Under range:

 A blinking 

“0”

 indicates that 

the sample absorbs less light than the zero 
reference. Check the procedure and make sure 
you use the same cuvette for reference (zero) 
and measurement.

Over Range:

 A flashing value of the maximum 

concentration indicates an over range condition. 
The concentration of the sample is beyond the 
programmed range: dilute the sample and 
re-run the test.

Standard Low:

 The standard reading is less 

than expected.

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the LCD 

displays dashes and “P1” (Calcium) and “P2” 

(Magnesium) the meter is ready. The code that 

appears on the secondary display is the one of 

the last selected parameter. If necessary, press

 

RANGE/GLP

 to change parameter. The 

blinking 

“ZERO”

 indicates that the instrument 

needs to be zeroed first.

Cap error:

 Appears when external light enters 

in the analysis cell. Assure that the cuvette cap 
is present.

Summary of Contents for HI96752

Page 1: ...INSTRUCTION MANUAL HI96752 Calcium and Magnesium High Range ISM ...

Page 2: ...port contact your local Hanna Instruments Office or e mail us at tech hannainst com Find your local Hanna Instruments Office at www hannainst com PRELIMINARY EXAMINATION For more details about spare parts and accessories see Accessories SPECIFICATIONS Range Calcium 0 to 400 mg L Magnesium 0 to 150 mg L Resolution 1 mg L Accuracy 25 C 77 F Calcium 10 mg L 5 of reading Magnesium 3 mg L 3 of reading ...

Page 3: ... measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state of the battery 4 The hourglass appears when an internal check is in progress 5 Status messages 6 The chronometer appears when the reaction timer is running 7 The month day and date icons appear when a date is displayed 8 ...

Page 4: ...Oxalate reagent Magnesium Using another 1 mL syringe add to the cuvette exactly 0 5 mL of sample Note Do not mix up the two 1 mL syringes 9 Calcium Replace the cap and invert the cuvette 10 times to mix Magnesium Replace the cap and invert several times to mix 10 Replace the cuvette into the holder and en sure that the notch on the cap is positioned securely into the groove 11 Press READ TIMER for...

Page 5: ... measurement Please check the preparation of the zero cuvette Light High There is too much light to perform a measurement Please check the preparation of the zero cuvette Inverted cuvettes The sample and the zero cuvette are inverted Zero A zero reading was not taken Follow the instructions of the measurement procedure for zeroing the meter Under range A blinking 0 indicates that the sample absorb...

Page 6: ...n 14 days from the date of purchase WARRANTY IST96752 01 18 Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Reagent Sets HI93752 01 Reagents for 1...

Page 7: ...the measurement phase 5 After a few seconds the display will show 0 0 The meter is now zeroed and ready for validation 6 Remove the cuvette 7 Place the specific CAL CheckTM Standard Cuvette B into the cuvette holder for Calcium B HI96752 11 Magnesium B HI96754 11 Ensure that the notch on the cap is positioned securely into the groove 8 Press CAL CHECK key and the lamp cuvette and detector icons to...

Page 8: ...blinking READ asks for reading calibration standard 8 Remove the cuvette 9 Place the specific CAL CheckTM Standard Cuvette B into the cuvette holder for Calcium B HI96752 11 Magnesium B HI96754 11 Ensure that the notch on the cap is positioned securely into the groove 10 Press READ TIMER and the lamp cuvette and detector icons will appear on the display depending on the measurement phase 11 The in...

Page 9: ...linking 17 Press RANGE GLP to edit the desired day 01 31 If the key is kept pressed the day number is automatically increased Note It is possible to change the editing from day to year and to month by pressing READ TIMER 18 Press ZERO CFM to save the calibration date 19 The instrument displays Stor for one second and the calibration is saved 20 The instrument will return automatically to measureme...

Page 10: ... empty and accurate measurements can t be taken any more the instrument shows dEAd bAtt and turns off To restart the instrument the battery must be replaced with a fresh one To replace the instrument s battery follow the steps Turn the instrument off by pressing ON OFF Turn the instrument upside down and remove the battery cover by turning it counterclockwise Extract the battery from its location ...