12
Under range:
A blinking
“0.00”
indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range:
A flashing value of the maximum concentration
indicates an over range condition. The concentration of the sample
is beyond the programmed range: dilute the sample and re‑run
the test.
c) during calibration procedure
Standard Low:
The standard reading is less than expected.
Standard High:
The standard reading is higher than expected.
Cap error:
Appears when external light enters in the analysis cell.
Assure that the cuvette cap is present.
Cooling lamp:
The instrument waits for the lamp to cool down.
d) other errors and warnings
ERRORS AND W
ARNINGS