 is an auto diagnostic portable microprocessor meter that benefits from Hanna 

Instruments’ years of experience as a manufacturer of analytical instruments. It has the advanced 
optical system based on a special Tungsten Lamp and a narrow band interference filter that allows 
most accurate and repeatable readings. All instruments are factory calibrated and the electronic 
and optical design minimizes the need of frequent calibration.
With the powerful 

CAL Check

TM

 validation function, you are able to validate good performance 

of your instrument at any time. The validation procedure is extremely user friendly. Just use the 
exclusive Hanna Instruments ready‑made, NIST traceable standards to verify the performance of 
the instrument and recalibrate if necessary.
All instruments are splash proof and the lamp and filter units are protected from dust or dirt by 
a transparent cup. This makes the instruments fulfill field applications. Display messages aid the 
user in routine operation. The meter has an auto‑shut off feature that will turn off the instrument 
after 10 minutes of non use in measurement mode or after 1 hour if left in calibration mode.
The meter uses an exclusive positive‑locking system to ensure that the cuvette is in the same 
position every time it is placed into the measurement cell. It is designed to fit a cuvette with a 
larger neck making it easier to add both sample and reagents. The cuvette is made from special 
optical glass to obtain best results.
The 

HI96707

 meter measures the Nitrogen‑Nitrite (N0

2

‑N) content in the samples in the 0.000 

to 0.600mg/L (ppm) range, in drinking, surface and saline water samples and in domestic and 
industrial wastes. The method is an adaptation of an EPA approved method.
The reagent is in

 

powder form and it is supplied in packets. The amount of reagent is precisely 

dosed to ensure the maximum repeatability.

degree Celsius 
degree Fahrenheit
milligrams per liter. mg/L is equivalent to ppm (parts per million)
milliliter
millivolt
American Society for Testing and Materials

°C

:

°F

:

mg/L

:

mL

:

mV

:

ASTM

:

GENERAL DESCRIPTION

ABBREVIA

TIONS

Summary of Contents for HI96707

Page 1: ...INSTRUCTION MANUAL HI96707C Nitrite Low Range ISM...

Page 2: ...the instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our worldwide contact list at www ha...

Page 3: ...PRINCIPLE OF OPERATION FUNCTIONAL DESCRIPTION ERRORS AND WARNINGS GENERAL TIPS FOR AN ACCURATE MEASUREMENT STARTUP MEASUREMENT PROCEDURE VALIDATION PROCEDURE CALIBRATION PROCEDURE GLP BATTERY MANAGEM...

Page 4: ...Meter is supplied complete with Sample Cuvettes and Caps 2 pcs CAL CheckTM standard cuvettes 9V Battery 9V Battery Scissors Cloth for wiping cuvettes Instrument quality certificate Instruction Manual...

Page 5: ...splay messages aid the user in routine operation The meter has an auto shut off feature that will turn off the instrument after 10 minutes of non use in measurement mode or after 1 hour if left in cal...

Page 6: ...non use in calibration mode with last reading reminder Dimensions 192 x 104 x 69 mm 7 6 x 4 1 x 2 7 Weight 320 g 11 3 oz REQUIRED REAGENTS Code Description Quantity test HI93707 0 Nitrite Low Range R...

Page 7: ...centration of the substance d optical path through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors are known Photometric chemica...

Page 8: ...ment process is carried out in two phases first the meter is zeroed and then the actual measurement is performed The cuvette has a very important role because it is an optical element and thus require...

Page 9: ...e press to confirm factory calibration restore READ TIMER this is a multi functional key In measurement mode press to make a measurement or press and hold for three seconds to start a pre programmed c...

Page 10: ...actory calibration 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge sta...

Page 11: ...check the preparation of the zero cuvette Light Low There is not enough light to perform a measurement Please check the preparation of the zero cuvette No Light The instrument cannot adjust the light...

Page 12: ...ge condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standard Low The standard reading is less than expected...

Page 13: ...of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet It is important that the sample does not cont...

Page 14: ...nently stained All the reaction times reported in this manual are at 25 C 77 F In general the reaction time should be increased for temperatures lower than 20 C 68 F and decreased for temperatures hig...

Page 15: ...n by pressing ON OFF The display briefly shows all tags on When the beeper sounds briefly and the LCD displays dashes the meter is ready The blinking ZERO indicates that the instrument needs to be zer...

Page 16: ...oned securely into the groove Press and hold READ TIMER for three seconds The display will show the countdown prior to measurement The beeper is playing a beep at the end of countdown period MEASUREME...

Page 17: ...pm ferric ions Fe3 above 2 ppm INTERFERENCES MEASUREMENT PROCEDURE At the end of measurement the instrument directly displays concentration in mg L of nitrogen nitrite NO2 N on the LCD To convert the...

Page 18: ...n please perform test at room temperature 18 to 25 C 64 5 to 77 0 F Turn the meter on by pressing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Place the CAL Che...

Page 19: ...ent Remove the cuvette Place the CAL CheckTM Standard HI96707 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press CAL CHECK and the l...

Page 20: ...ible to interrupt calibration procedure at any time by pressing CAL CHECK or ON OFF keys Warning Do not calibrate the instrument with standard solutions other than the Hanna Instruments CALCheckTM Sta...

Page 21: ...for reading calibration standard Remove the cuvette Place the CAL CheckTM Standard HI96707 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the gr...

Page 22: ...8 2009 appears on the display or 01 01 2009 if the factory calibration was selected before In both cases the year number is blinking ready for date input CALIBRATION PROCEDURE Press GLP to edit the de...

Page 23: ...ressed the day number is automatically increased Note It is possible to change the editing from day to year and to month by pressing READ TIMER Press ZERO CFM to save the calibration date The instrume...

Page 24: ...isplay LAST CALIBRATION DATE If no calibration was performed the factory calibration message F CAL will appear on the main display and the instrument returns to measurement mode after three seconds FA...

Page 25: ...t is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The remaining battery capacity is evaluate...

Page 26: ...resh one Insert back the battery cover and turn it clockwise to close BATTERY REPLACEMENT ACCESSORIES MAN96707 01 18 Reagent Set HI93707 01 Reagents for 100 tests HI93707 03 Reagents for 300 tests Oth...

Page 27: ...6707 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions This warranty is limited to repair or repla...

Page 28: ...a Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments USA 270 George Washington Highway Smithfield RI 02917 Phone 800...

Reviews: