20
Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK
or ON/OFF keys.
Warning
: Do not calibrate the instrument with standard solutions other than the Hanna Instruments
CAL Check
TM
Standards, otherwise erroneous results will be obtained.
• Press
ZERO/CFM
and the lamp, cuvette and detector icons will appear on the display, depending
on the measurement phase.
CALIBRA
TION
PROCEDURE
• Turn the meter on by pressing
ON/OFF
.
• When the beeper sounds briefly and the LCD displays dashes, the meter is ready.
• Press and hold
CAL CHECK
for three seconds to enter calibration mode. The display will show
“CAL”
during calibration procedure. The blinking
“ZERO”
asks for instrument zeroing.
• Place the
CAL Check
TM
Standard
HI96705-11
Cuvette A into the cuvette holder and ensure
that the notch on the cap is positioned securely into the groove.
3''