26
The
HI88703
has built in complete GLP information. The calibration date and the calibration points
are displayed in a comprehensive mode for each range.
To display the GLP information, simply press CAL key. A screen with instrument serial number and with
information about the calibration is displayed. For further information, press the GLP key.
The GLP contains:
• Instrument serial number
• The last user calibration date, in selected
format and time in hh.mm format. If no calibration was
performed, the “Factory Calibration” message is displayed.
• Parameter as Ratio Turbidity or Non Ratio
Turbidity.
• The value of each calibration point (up to
5 points for ratio turbidimeter and 2 points for non ratio
turbidimeter). If the first calibration point was skipped,
the 0.00 value is displayed.
GOOD LABORATORY PRACTICE (GLP)
To restore the factory calibration for the currently selected
range, press
CAL
key when in the main screen. The first
screen for the GLP is displayed. Press the
Delete
key to
initiate the calibration delete procedure and then press
the
CFM
key to delete the user calibration and restore the
Factory calibration.
Note:
Only the user calibration for the current selected
range is deleted.
RESTORE FACTORY CALIBRATION