•  Turn the meter on by pressing the button. All 

segments will be displayed. When the display 
shows “

Add

”, “

C.1

” with “

Press

” blinking, the 

meter is ready.

•  Fill the cuvette with 10 mL of unreacted sample 

and replace the cap. Place the cuvette into the 
meter and close the meter’s cap.

•  Press the button. When the display shows 

Add

”, “

C.2

” with “

Press

” blinking, the meter 

is zeroed.

Note: Any chlorine present in the sample 

will interfere with the reading. To remove 

the chlorine interference add one drop 

of HI93755-53 Chlorine Remover to the 

unreacted sample.

•  Remove the cuvette from the meter and unscrew 

the cap. Using a 1 mL syringe carefully add 
exactly 1 mL of 

HI775S

 Reagent to the 

sample. Replace the cap and gently invert  
5 times. Place the cuvette back into the meter.

Note: Pay attention not to spill reagent 

otherwise full color development may be 

inhibited.

•  Press the button. The instrument directly 

displays the concentration of alkalinity in ppm 
of CaCO

3

. Alkalinity conversion: 

 

1 ppm CaCO

3

 = 0.02 meq/L

 

 

= 0.056 dkH 

  The meter automatically turns off after  

2 minutes.

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you 
use the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us at 
[email protected]
Find your local Hanna Instruments Office on www.hannainst.com.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI775

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 25 tests
• 1 mL syringe with tip
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Range

0 to 500 ppm

Resolution

1 ppm

Accuracy

±5 ppm ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @610 nm

Light Detector

Silicon Photocell

Method

Colorimetric method. The reaction causes a distinctive range of colors 
from yellow to blue to develop. This meter has been developed to work 
with fresh water samples.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 10 minutes of non-use and 2 minutes after reading

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Functional Description

Errors and Warnings

Measurement Procedure