Measurement Procedure
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each
HI774
meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 10 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Thank You
Thank you for choosing a Hanna Instruments product. Please read this instruction manual
carefully before using the instrument.
For more information about Hanna Instruments and our products, visit www.hannainst.com.
For technical support, contact your local Hanna Instruments Office or e-mail us at
[email protected]
Preliminary Examination
Specifications
Range
0.00 to 0.90 ppm
Resolution
0.01 ppm
Accuracy
±0.02 ppm ±5% of reading @25 °C/ 77 °F
Light Source
Light Emitting Diode @525 nm
Light Detector
Silicon Photocell
Method
Adaptation of Standard Methods for the Examination of Water and
Wastewater, 20
th
edition, Ascorbic Acid method. The reaction be-
tween phosphate and the reagent causes a blue tint in the sample.
Environment
0 to 50 °C (32 to 122 °F); RH max 95% non-condensing
Battery Type
1.5V AAA (1 pc.)
Auto-Shut off
After 7 minutes of non-use and 2 minutes after reading
Dimensions
86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight
52 g (1.84 oz.)
Light High:
There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Light Low:
There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes:
The sample and the zero cuvette are inverted.
Under Range:
A blinking
“0.00”
indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you use
the same cuvette for reference (zero) and measurement.
Over Range:
A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low:
The battery must be replaced soon.
Dead Battery:
This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the meter.
Functional Description
Errors and Warnings
Cover
Cuvette with cap
Cuvette holder
LCD Liquid Crystal Display
Operation Button
• Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows “
Add
”, “
C.1
” with “
Press
” blinking,
the meter is ready.
• Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
• Press the button. When the display shows
“
Add
”, “
C.2
” with “
Press
” blinking the
meter is zeroed.
• Remove the cuvette from the meter and unscrew
the cap. Add the content of one packet of
HI774-0
reagent. Replace the cap and shake
gently for 2 minutes until the powder is
completely dissolved. Place the cuvette back
into the meter.
• Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement) or,
alternatively, wait for 3 minutes and press the
button.
• The instrument directly displays the
concentration of phosphate in ppm. The
meter automatically turns off after 2 minutes.