Hanna Instruments HI772 Instruction Manual Download Page 2

Turn the meter on by pressing the button. All segments 
will be displayed. When the display shows “

Add

”, 

C.1

” with “

Press

” blinking, the meter is ready.

Fill the cuvette with 10 mL of unreacted sample and 
replace the cap. Place the cuvette into the meter and 
close the meter’s cap.

Press the button. When the display shows “

Add

”, 

C.2

” with “

Press

” blinking the meter is zeroed.

Remove the cuvette from the meter and unscrew the 
cap. Using a 1 mL syringe carefully add exactly 1 mL 
of 

HI772S

 Reagent to the sample. Replace the cap 

and gently invert 5 times. Place the cuvette back 
into the meter.

Note: Pay attention not to spill reagent 

otherwise full color development may be 

inhibited.

Press the button. The instrument directly displays 
the concentration of alkalinity in dKH. Alkalinity 
conversion: 
1 dKH 

= 17.86 ppm CaCO

3

 

= 0.358 meq/L

The meter automatically turns off after 10 minutes.

Measurement Procedure

HI772S

Thank You

Thank you for choosing a Hanna Instruments product. Please read this 
instruction manual carefully before using the instrument.  

For more information about Hanna Instruments and our products, visit  
www.hannainst.com.

For technical support, contact your local Hanna Instruments Office or e-mail us 
at  [email protected]

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI772

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 25 tests
• 1 mL syringe with tip
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0.0”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you use 
the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Functional Description

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Errors and Warnings

Range

0.0 to 20.0 dKH

Resolution

0.1 dKH

Accuracy

±0.3 dKH ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @610 nm

Light Detector

Silicon Photocell

Method

Colorimetric method. The reaction causes a distinctive range of 
colors from yellow to green to greenish blue to develop.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 10 minutes of non-use

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications

Hanna Instruments reserves the right to modify the design, construction or appearance of 

its products without advance notice.