Hanna Instruments HI770 Instruction Manual Download Page 2

•  Fill both cuvettes with 10 mL of unreacted 

sample. Replace the cap on one of the cuvettes, 
this is the zero.

•  To the second cuvette add one packet of 

HI770A-25

 reagent. Replace the cap and shake 

gently until the reagent is completely dissolved.

 

Unscrew the cap and add one packet of 

HI770B-25

 

reagent. Replace the cap and 

shake vigorously until the reagent is completely 
dissolved. Wait 10 minutes.

•  Unscrew the cap and add one packet of 

HI770C-25

 reagent. Replace the cap and shake 

gently until the reagent is completely dissolved. 
Wait 2 minutes.

•  During the two minute reaction time. Turn the 

meter on by pressing the button. All segments 
will be displayed. When the display shows 

Add

”, “

C.1

” with “

Press

” blinking, the meter 

is ready.

•  Place the first cuvette (unreacted) into the meter 

and close the meter’s cap. Press the button. 
When the display shows “

Add

”, “

C.2

” with 

Press

” blinking the meter is zeroed.

•  After the two minute reaction time, place the 

second cuvette (reacted) into the meter and 
close the meter’s cap. Press the button.

•  The instrument directly displays the concentration 

in ppm of silica on the LCD. To convert the 
reading to ppm silicon (Si) multiply the reading 
by 0.467. The meter automatically turns off after 
2 minutes.

Measurement Procedure

Hanna Instruments reserves the right to modify the design, construction or appearance of 

its products without advance notice.

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you 
use the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Functional Description

Errors and Warnings

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us at 
[email protected]
Find your local Hanna Instruments Office on www.hannainst.com.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI770

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 6 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Range

0 to 200 ppm

Resolution

1 ppm

Accuracy

±2 ppm ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @470 nm

Light Detector

Silicon Photocell

Method

Adaptation of the USEPA method 370.1 and Standard Method 
4500-SiO

2

 C.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 10 minutes of non-use and 2 minutes after reading

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications