•  Turn the meter on by pressing the button. All 

segments will be displayed. When the display 

shows “

Add

”, “

C.1

” with “

Press

” blinking, the 

meter is ready.

•  Fill one cuvette up to the 10 mL mark with 

deionized water and replace the cap. This is 

the blank.

•  Place the blank cuvette into the meter and close 

the meter’s cap.

•  Press the button. When the display shows 

Add

”, “

C.2

” with “

Press

” blinking the meter 

is zeroed.

•  Fill second cuvette up to the 10 mL mark with 

unfiltered sample and replace the cap. This is 

the apparent color.

•  Insert the apparent color cuvette into the meter 

and close the meter’s cap.

•  Press the button and the meter directly displays 

the value of apparent color.

•  Remove the cuvette and press the button twice 

to restart the meter.

•  Filter 50 mL of sample through a 0.45 

µ

membrane filter into a beaker.

•  After measuring of apparent color rinse the 

cuvette three times with filtered sample and 

then fill it up to the 10 mL mark with the 

filtered sample and replace the cap. This is 

the true color.

•  Follow the procedure above to re-zero the meter 

using the blank cuvette.

•  Insert the true color cuvette into the instrument.
•  Press the button and the meter directly displays 

concentration in color units of the true color on 

the LCD. The meter automatically turns off after 

2 minutes.

Measurement Procedure

Hanna Instruments reserves the right to modify the design, construction or appearance of 

its products without advance notice.

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you 
use the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Functional Description

Errors and Warnings

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us at 
[email protected]
Find your local Hanna Instruments Office on www.hannainst.com.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI727

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Range

0 to 500 g/L PCU

Resolution

5 PCU

Accuracy

±10 PCU ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @470 nm

Light Detector

Silicon Photocell

Method

Adaptation of the Standard Methods for the Examination of Water 
and Wastewater 21

th

 edition, Colorimetric Platinum Cobalt method.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 10 minutes of non-use and 2 minutes after reading

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications