•  Turn the meter on by pressing the button. All 

segments will be displayed. When the display 
shows “

Add

”, “

C.1

” with “

Press

” blinking, the 

meter is ready.

•  Fill the cuvette with 10 mL of unreacted sample 

and replace the cap. Place the cuvette into the 
meter and close the meter’s cap.

•  Press the button. When the display shows 

Add

”, “

C.2

” with “

Press

” blinking the meter 

is zeroed.

•  Remove the cuvette from the meter and unscrew 

the cap. Add 4 drops of 

HI715A-0

 reagent A. 

Replace the cap and swirl the solution.

•  Unscrew the cap and add 4 drops of 

HI715B-0 

reagent B. Replace the cap and swirl the 
solution. Place the cuvette back into the meter.

•  Press and hold the button until the timer is 

displayed on the LCD (the display will show 
the countdown prior to the measurement) 
or, alternatively, wait for 3 minutes and 
30 seconds and press the button.

•  The instrument displays the results in mg/L 

(ppm) of ammonia nitrogen (NH

3

-N). To convert 

the reading to ppm of ammonia (NH

3

), multiply 

the reading by factor of 1.214. The meter 
automatically turns off after 10 minutes.

Measurement Procedure

Hanna Instruments reserves the right to modify the design, construction or appearance of 

its products without advance notice.

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Functional Description

Errors and Warnings

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0.00”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you use 
the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us at 
[email protected]
Find your local Hanna Instruments Office on www.hannainst.com.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI715

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 25 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Range

0.00 to 9.99 ppm NH

3

-N

Resolution

0.01 ppm

Accuracy

±0.05 ppm ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @470 nm

Light Detector

Silicon Photocell

Method

Adaptation of the ASTM Manual of Water and Environmental 
Technology, D1426, Nessler Method. The reaction between ammo-
nia and reagents causes a yellow tint in the sample.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 10 minutes of non-use

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications