17
ACCESSORIES
Code and
serial number
on the other
side
Temperature
sensor
Quick connector with
Rubber protection, and
1 meter cable
Reference
junction
pH bulb
ELECTRODE
MAINTENANCE
PREPARATION
• Remove the protective cap. DO NOT BE ALARMED IF ANY
SALT DEPOSITS ARE PRESENT. Rinse with water.
• Shake the electrode down as you would do with a clini‑
cal thermometer to eliminate any air bubbles inside the
glass bulb.
• Remove the fill hole cover to ensure the reference junctions
are flowing. Set aside for storage.
• Top off the electrolyte filling solution using
HI7082
solution.
• If the bulb and/or junction are dry, soak the electrode
in
HI70300
Storage Solution for at least 30 minutes.
• Rinse with water.
• Calibrate before using.
Loss of shielding due to low electrolyte level:
Empty electrolyte with a syringe and refill with fresh
HI7082.
STORAGE
• To ensure a quick response, the glass bulb and the junction
should be kept moist and not allowed to dry.