Remove the instrument from the packing material and examine 

it carefully. If any damage has occurred during shipment, 

immediately contact your local Hanna Instruments Office.
The meter is supplied with:
• 

HI1293D

 pH-electrode

• temperature probe (fixed)
• pH4.01 and pH7.01 buffer solutions (20 mL each)
• 12 Vdc power adapter and instructions
• quality certificate

Note: Conserve all packing material until the instrument 

has been observed to function correctly. Any defective 

item must be returned in its original packing. 

PRELIMINARY EXAMINATION

HI991401 

indicator features a large, dual-level backlit LCD to 

give istantaneous readings of both pH and temperature, even 

from a distance.
This waterproof indicator has been specially designed to meet 

the grower’s need for equipment suited to the aggressive 

environments of agricultural and hydroponics applications.
Measurements are highly accurate and the meter can be cali-

brated at one or two points. Low battery and battery changes 

are no longer a problem thanks to a 12 Vdc power supply 

included with the meter.
The 

HI1293D

 pH electrode & the temperature probe have been 

molded with pipe thread (1/2’’ NPT), which allows the user to 
attach them to an in-line system. 
Moreover, the pH electrode has been specially designed to 

guarantee greater clogging resistance in fertilizer solutions with 

high concentrations of phosphate and nitrate.

GENERAL DESCRIPTION

FUNCTIONAL DESCRIPTION

SPECIFICATIONS

  1.  Liquid Crystal Display

  2.  ON/OFF/MODE button

  3.  Power supply connector

  4.  12 Vdc power adapter

 5. 

HI1293D 

pH electrode with differential input and pipe 

thread 1/2’’ NPT

  6.  Temperature probe with pipe thread 1/2’’ NPT 

    (also working as matching pin for pH electrode)

  7.  DIN connector

  8.  SET/HOLD button

Dear Customer,
Thank you for choosing a Hanna Instruments product. 
This manual will provide you with the necessary information 
for a correct operation. Please read it carefully before using 
the meter. If you need additional technical information, do not 
hesitate to e-mail us at [email protected].
This instrument is in compliance with the      

directives.

Instruction Manual

HI991401

Range

0.0 to 14.0 pH
0.0 to 60 ºC / 32.0 to 140 ºF

Resolution

0.1pH / 0.1 ºC / 0.1 ºF

Accuracy
@20 ºC/68 ºF

±0.1 pH / ±0.5 ºC / ±1 ºF

Typical EMC 
Deviation

±0.1 pH / ±1 ºC / ±2 ºF

pH Calibration

Automatic, 1 or 2 point
with auto-buffer recognition

Temperature 
Compensation

Automatic

Probes

HI1293D

 pH electrode (included)

HI1294 

temperature probe (fixed)

Power supply

12 Vdc power adapter (included)

Environment

0 to 50 ºC (32 to 122 ºF);
RH max 95% non-condensing

Dimensions

160 x 105 x 31 mm (6.2 x 4.1 x 1.2’’)

Weight

190 g (6.7 oz.)

OPERATIONAL GUIDE

Preliminary operations

Connect the pH electrode to the meter.
Connect the 12 Vdc adapter to the meter and to the mains; 
the display will lit. 
Press and hold the MODE button for 2-3 seconds. All the used 
segments on the LCD will be visible for a few seconds.

To change the temperature unit

To change the temperature unit (from °C to °F), from measure-
ment mode, press and hold the MODE button until “TEMP” 
and the current temperature unit are displayed on the lower 
LCD (E.g. TEMP °C).
Use the SET/HOLD button to change the temperature unit, and 
then press MODE button twice to return to normal measurement 
mode.

To freeze the display

Press and hold the SET/HOLD button for 2-3 seconds, until 
“HOLD” appears on the secondary display. 
Press either button to return to normal mode.

www.hannainst.com

Summary of Contents for HI 991401

Page 1: ...Instruction Manual HI991401 www hannainst com ...

Page 2: ...uited to the aggressive environments of agricultural and hydroponics applications Measurements are highly accurate and the meter can be cali brated at one or two points Low battery and battery changes are no longer a problem thanks to a 12 Vdc power supply included with the meter The HI1293D pH electrode the temperature probe have been molded with pipe thread 1 2 NPT which allows the user to attac...

Page 3: ...F MODE button 3 Power supply connector 4 12 Vdc power adapter 5 HI1293D pH electrode with differential input and pipe thread 1 2 NPT 6 Temperature probe with pipe thread 1 2 NPT also working as matching pin for pH electrode 7 DIN connector 8 SET HOLD button ...

Page 4: ...y operations Connect the pH electrode to the meter Connect the 12 Vdc adapter to the meter and to the mains the display will lit Press and hold the MODE button for 2 3 seconds All the used segments on the LCD will be visible for a few seconds To change the temperature unit To change the temperature unit from C to F from measure ment mode press and hold the MODE button until TEMP and the current te...

Page 5: ...any measurement make sure the meter has been calibrated To clear a previous calibration press the MODE button after entering the calibration mode The lower LCD will display ESC for 1 second and the meter will return to normal measurement mode The CAL symbol on the LCD will disappear The meter will be reset to the default calibration If measurements are taken in different samples successively rinse...

Page 6: ...isplay pH7 01 USE or pH6 86 USE if you have selected the NIST buffer set The CAL tag blinks on the display For a single point pH calibration immerse the electrode and the temperature probe in any buffer from the selected buffer set eg pH7 01 or pH4 01 or pH10 01 The meter will recognize the buffer value automatically If using pH4 01 or pH10 01 the meter will display OK for 1 second and then return...

Page 7: ...het 25 pcs HI70010P pH10 01 solution 20 mL sachet 25 pcs HI7004M pH4 01 solution 230 mL bottle HI7006M pH6 86 solution 230 mL bottle HI7007M pH7 01 solution 230 mL bottle HI7009M pH9 18 solution 230 mL bottle HI7010M pH10 01 solution 230 mL bottle HI7061M Electrode cleaning solution 230 mL bottle HI70300M Electrode storage solution 230 mL bottle HI710005 12 Vdc power adapter US plug HI710006 12 Vd...

Page 8: ...e returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner IST991401 01 19 RECOMMENDA...

Reviews: