To connect the probe

With the meter turned off, connect the 

HI  1296D

probe to the DIN socket on the top of the meter by

aligning the pins and pushing in the plug. Tighten

the nut to ensure a good connection. Remove

the protective cap from the probe before taking

any measurements.

To turn the meter ON and check the bat-

tery status

Press the ON/OFF/MODE button until the dis-

play lights up. At start-up, all the LCD segment s

are displayed for 1 second, then the percent in-

dication of the remaining battery life is shown for

another second (E.g. % 100 BA TT). The meter

then enters the normal measuring mode.

Note

: If the display needs to be checked, keep

the ON button pressed while turning the meter

on. The meter will display all segments as

long as the button is pressed.

To freeze the display

While in measurement mode, press the SET/

HOLD button, HOLD appears on the secondary

display and the reading will be frozen on the LCD

(E.g. pH 5.73 HOLD). Press any button to return

to normal mode.

To turn the meter OFF

While in normal measurement mode, press the

ON/OFF/MODE button. OFF will appear on the

secondary display. Release the button.

Note

: The meter is provided with an acoustic

signal feature, which can be disabled using

the switch located in the battery compartment.

Note

: When the meter detects the absence of a

temperature probe at it s input, the  Automatic

Temperature Compensation is turned off, and

the meter uses a default value of 25°C (77°F)

for the temperature measurement and com-

pensation. In this condition, the secondary

LCD shows 25.0°C (77.0°F) blinking. When a

probe is connected, the meter automatically

returns to the ATC mode, the ATC tag is turned

on, and the temperature is shown on the sec-

ondary display.

OPERATIONAL GUIDE

METER SETUP

pH MEASUREMENT & CALIBRATION

BATTERY REPLACEMENT

Setup mode allows the selection of the tempera-

ture unit and of the pH buffer set.

To enter the Setup mode, press and hold the

ON/OFF/MODE button until CAL on the second-

ary display is replaced by TEMP and the current

temperature unit (E.g. TEMP °C). Then:

for °C/°F selection

, use the SET/HOLD button.

After the temperature unit has been selected,

press ON/OFF/MODE to enter the buffer set

selection mode; press ON/OFF/MODE twice

to return to the normal measuring mode.

to change the calibration buffer set

, after set-

ting the temperature unit, the meter will show

the current buffer set: “pH 7.01 BUFF” (for 4.01/

7.01/10.01) or “pH 6.86 BUFF” (for 4.01/6.86/

9.18). Change the set with the SET/HOLD but-

ton, then press ON/OFF/MODE to return to nor-

mal mode.

• Make sure the meter has been calibrated be-

fore use.

• If the probe is dry, soak it in 

HI 70300

 storage

solution for one hour to reactivate it.

• Place the tip of the probe into the sample to be

tested.

• Stir briefly and wait until the   symbol on the

top left of the LCD is turned off.

• The LCD shows the pH value (automatically

compensated for temperature) on the primary

LCD, while the secondary LCD shows the tem-

perature of the solution.

• If measurements are taken in different

samples successively, rinse the probe tip thor-

oughly to eliminate cross-cont amination. After

cleaning, rinse the probe tip with some of the

sample to be measured.

pH calibration

For better accuracy, frequent calibration of the

instrument is recommended. In addition, the

instrument must be recalibrated whenever:

a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where high accuracy is required.
d) At least once a month.

• From normal mode, press and hold ON/OFF/

MODE until OFF on the secondary display is

replaced by CAL. Release the button.

• The LCD enters the calibration mode, display-

ing “pH 7.01 USE” (or “pH 6.86 USE” if the

NIST buffer set was selected). After 1 second

the meter activates the automatic buffer rec-

ognition feature. If a valid buffer is detected,

then its value is shown on the primary display,

and REC appears on the secondary LCD. If

no valid buffer is detected, the meter keeps

the USE indication active for 12 seconds, and

then replaces it with WRNG, indicating that the

sample being measured is not a valid buf fer.

• For a 

single-point calibration

 with buffers pH

4.01, 9.18 or 10.01, the meter automatically

accepts the calibration when the reading is

stable; the meter will show on the primary dis-

play the accepted buffer , with the message

“OK 1” on the secondary display , and an au-

dible signal is produced.

After 1 second the meter automatically returns

to the normal measuring mode.

If a single-point calibration with buffers pH 7.01

or 6.86 is desired, then after the calibration

point has been accepted the ON/OFF/MODE

button must be pressed in order to return to

the normal measuring mode.  After the button

is pressed, the meter shows "7.01" (or "6.86")

- "OK 1", and an audible signal is produced.

After 1 second, the meter automatically returns

to the normal measuring mode.

Note

: It is always recommended to carry out a

two-point calibration for better accuracy.

• For a 

two-point calibration

, place the probe in

pH 7.01 (or pH 6.86) buffer. After the calibration

point has been accepted, the “pH 4.01 USE”

message appears. The message is held for

12 seconds, unless a valid buffer is recog-

nized. If no valid buffer is recognized, then the

WRNG message is shown. If a valid buffer

(pH 4.01,  pH 10.01 or pH 9.18) is detected,

then the meter completes the calibration pro-

cedure. When the buffer is accepted, the LCD

shows the accepted value with the “OK 2”

message on the secondary display. The meter

then returns to the normal measuring mode.

Note

: When the calibration procedure is com-

pleted, the CAL tag is turned on.

To quit calibration and to reset to the de-

fault values

• After entering the calibration mode and before

the first point is accepted, it is possible to quit

the procedure and return to the last calibration

data by pressing the ON/OFF/MODE button.

The secondary LCD displays ESC for 1 sec-

ond and the meter returns to normal mode.

• To reset to the default values and clear a pre-

vious calibration, press the SET/HOLD button

after entering the calibration mode and before

the first point is accepted.

The secondary LCD displays CLR for 1 sec-

ond, the meter resets to the default calibration

and the “CAL” tag on the LCD disappears.

The meter displays the remaining battery

percentage when turned on. When the level

is below 5%, the 

- +

 symbol on the bottom

left of the LCD blinks to indicate a low battery

condition. If the battery level is low enough to

cause erroneous readings, the Battery Error

Prevention System (BEPS) turns the meter

off.

Unscrew the 4 screws located on the back

of the meter and carefully replace the 3  AA

batteries located in the battery compartment,

while paying attention to their polarity. Reat-

tach the back making sure that the gasket is

in place and tighten the 4 screws to ensure

a watertight seal.