background image

13

12

The setup mode is used to view or change the instrument parameters.

To enter setup press ALT+FNC when the
meter is in measurement mode.

"Set" is displayed on the upper LCD. The
lower LCD displays the blinking code of the
current setup item.

Select the desired setup item using
the 

 or 

 key.

Press ALT+CFM to confirm.

Note:

 If ALT+FNC are pressed before item confirmation, the meter

will escape and return to measurement mode.

Once the desired setup item has been
selected, its current value blinks (if it is a
changeable parameter).

To change the value use the 

 or

 key.

Press ALT+CFM to confirm.

Note:

 Press ALT+FNC before confirmation to escape without chang-

ing the previously set value.

SETUP

The conductivity of an aqueous solution is the measure of its ability to
carry an electrical current by means of ionic motion.
The conductivity invariably increases with increasing temperature.
It is affected by the type and number of ions in the solution and by
the viscosity of the solution itself. Both parameters are temperature
dependent. The dependency of conductivity on temperature is ex-
pressed as a relative change per degree Celsius at a particular
temperature, commonly as percent per °C.
The following table lists the temperature dependence of the HANNA
instruments

®

 calibration solutions.

°C

°F

HI 7030

HI 7031

HI  7033

HI  7034

HI 7035

HI 7039

HI 8030

HI 8031

HI 8033

HI 8034

HI 8035

HI 8039

(μS/cm)

(μS/cm)

(μS/cm)

(μS/cm)

(μS/cm)

(μS/cm)

0

32

7150

776

64

48300

65400

2760

5

41

8220

896

65

53500

74100

3180

10

50

9330

1020

67

59600

83200

3615

15

59

10480

1147

68

65400

92500

4063

16 60.8

10720

1173

70

67200

94400

4155

17 62.6

10950

1199

71

68500

96300

4245

18 64.4

11190

1225

73

69800

98200

4337

19 66.2

11430

1251

74

71300

100200

4429

20

68

11670

1278

76

72400

102100

4523

21 69.8

11910

1305

78

74000

104000

4617

22 71.6

12150

1332

79

75200

105900

4711

23 73.4

12390

1359

81

76500

107900

4805

24 75.2

12640

1386

82

78300

109800

4902

25

77

12880

1413

84

80000

111800

5000

26 78.8

13130

1440

86

81300

113800

5096

27 80.6

13370

1467

87

83000

115700

5190

28 82.4

13620

1494

89

84900

117700

5286

29 84.2

13870

1521

90

86300

119700

5383

30

86

14120

1548

92

88200

121800

5479

31 87.8

14370

1575

94

90000

123900

5575

CONDUCTIVITY VS TEMPERATURE CHART

Summary of Contents for HI 9835

Page 1: ...Instruction Manual HI 9835 Waterproof EC TDS NaCl C Meter for Lab and Field 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ...led to measure the actual conductivity The housing is completely waterproof and built to stand the harsh conditions of field use For extended time operations this meter can be connected to an external 12 Vdc power supply PRELIMINARY EXAMINATION GENERAL DESCRIPTION Dear Customer Thank you for choosing a HANNA instruments product Please read this instruction manual carefully before using the instru ...

Page 3: ... L NaCl 0 1 Temp 0 1 C Accuracy EC 1 of reading 0 05 μS cm or 1 digit TDS 1 of reading 0 03 ppm or 1 digit NaCl 1 of reading Temp 0 4 C Typical EMC EC 1 of reading Deviation TDS 1 of reading NaCl 1 of reading Temp 0 1 C FUNCTIONAL DESCRIPTION 1 Power adapter socket 2 Probe connector 3 Liquid Crystal Display LCD 4 ON OFF key to turn the meter on and off 5 ALT key to activate alternate key function ...

Page 4: ...ution to be tested The sleeve holes must be completely submerged Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve If needed press the RANGE key repeatedly until the desired range EC TDS or NaCl is selected on the LCD Allow for the reading to stabilize The upper LCD displays the measure in the selected range while the temperature is displayed on the lower LCD...

Page 5: ... Three options of compensating temperature are available Automatic Atc The probe has a built in temperature sensor the value of the temperature is used to automatically compensate the EC TDS reading using 25 C as reference temperature This is the default option Manual Mtc The temperature value shown on the lower LCD can be manually set by the user with the up and down arrow keys The compensation i...

Page 6: ...he NaCl range and press the CAL key Rinse the probe with some of the calibration solu tion or deionized water then immerse it in HI 7037 solution The sleeve holes must be completely sub merged Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve The BUF and CAL indications are dis played The upper LCD shows the uncalibrated NaCl reading in percentage The lower L...

Page 7: ...rature dependent The dependency of conductivity on temperature is ex pressed as a relative change per degree Celsius at a particular temperature commonly as percent per C The following table lists the temperature dependence of the HANNA instruments calibration solutions C F HI 7030 HI 7031 HI 7033 HI 7034 HI7035 HI 7039 HI8030 HI8031 HI8033 HI8034 HI8035 HI8039 μS cm μS cm μS cm μS cm μS cm μS cm ...

Page 8: ...the rear cover of the instrument and replace the four 1 5V AA bat teries with new ones paying attention to the correct polarity New batteries allow approximately 150 hours of continuous use A 12 Vdc power adapter can also be used It is recommended to use the HANNA volt age adapters that provide the proper polarity configuration However other adapters can be used In this case check the polarity of ...

Page 9: ...0L 12880 μS cm solution 500 mL bottle HI 7030M 12880 μS cm solution 230 mL bottle HI 70031P 1413 μS cm solution 20 mL sachet 25 pcs HI 7031L 1413 μS cm solution 500 mL bottle HI 7031M 1413 μS cm solution 230 mL bottle HI 70033P 84 μS cm solution 20 mL sachet 25 pcs HI 7033L 84 μS cm solution 500 mL bottle HI 7033M 84 μS cm solution 230 mL bottle HI 7034L 80000 μS cm solution 500 mL bottle HI 7034M...

Page 10: ...When shipping any instrument make sure it is properly packaged for complete protection Hanna Instruments reserves the right to modify the design construction and appearance of its products without advance notice Recommendations for Users Before using this product make sure that it is entirely suitable for the environment in which it is used Operation of this instrument in residential area could ca...