OPERATIONAL GUIDE

pH and ORP Electrodes Connection

In order to protect the instrument against vapors and humid-
ity, the BNC connectors are shielded behind waterproof sheaths.

• Slide the protective sheaths down. Connect the pH and

the ORP electrodes to the BNC connectors and then slide
the protective sheaths back up to the casing. For
maximum waterproof protection, make sure the connec-
tors are completely covered.

• Do not be alarmed if white crystals appear around the

electrode protective cap. This is normal with pH and
ORP electrodes and they dissolve when rinsed with
water.

Taking Measurements

• Turn the meter on by connecting the 12VDC power

adapter to the meter and the mains.

• Remove the protective cap from  the pH and ORP

electrodes.

• Immerse the tips (4cm/1½”)

of pH, ORP electrodes and
ground probe in the sample.
In order not to affect the
accuracy of measurements,
electrodes  should not touch
or stand close to the vessel’s
walls or bottom.

• The 2 LCDs will show the

pH and ORP values. Allow
the readings to stabilize.

Adjusting the Setpoint

With 

HI981406

 

you can select your own setpoints and be

alerted through two independent pH and ORP alarm LEDs
when an abnormal situation arises.

• Unscrew and remove the rear panel and gasket seal to

access the MEASURE/SET switches and move them to the
left (SET).

• With a small screwdriver adjust the setpoint trimmers to

display the desired value shown on the LCD. You can
choose to adjust one or both setpoints at once inside the
following ranges:
pH:

from 5.0 to 9.0

ORP:

from 350 to 850 mV

The pH LED blinks when the pH is higher than the
setpoint.
The ORP LED blinks when the ORP is lower than the
setpoint.

• Move the switches back to MEASURE and secure the rear

panel properly replacing the gasket.

Electrode Maintenance

• When not in use, rinse the electrodes with water to

minimize contamination and store them with a few
drops of storage (

HI 70300

) solution in the protective

cap. Always replace the protective cap when not in use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.

• If the electrodes have been left dry, soak the tips in a

storage (

HI 70300

) solution overnight to reactivate

them.

• To minimize clogging and provide longer life for the

electrodes, it is recommended to clean them at least once
a month. Immerse the tip of the electrodes in a 

HI 7061

cleaning solution

 

for half an hour and then rinse with

tap water.

ORP Electrode

Check the accuracy of the meter by immersing the ORP electrode
in a 

HI 7020

 

solution. The reading should be between 200

and 275 mV at 20°C (68°F). Otherwise clean the platinum tip
of the electrode by rubbing it with a soft cotton or lint-free cloth
soaked with 

HI 7061

 

cleaning solution, or alcohol

After cleaning, condition the ORP electrode by leaving it in a
pretreatment oxidizing (

HI 7092

) solution for half an hour.

The instrument must be recalibrated whenever:

a) The pH electrode is cleaned or replaced.

b) Where high accuracy is required.

c) At least once a month.

pH Calibration

Pour a small quantities of pH 7.0 (

HI 7007

) and pH 10.0

(

HI 7010

) solution (or alternatively pH4.0) into two clean

beakers.

For accurate calibration use two beakers for each buffer
solution, the first one for rinsing the tip of the electrode and
the second one for calibration.
pH 4.0 (

HI 7004

) is recommended for measuring acidic samples.

Use pH 10.0 (

HI 7010

) if subsequent samples are alkaline.

• Turn the meter on and make sure that the MEASURE/SET

switch is on MEASURE (inside the back cover).

• Remove the protective cap from the electrode, rinse and

immerse it in the pH 7.0 buffer. Stir gently and then
wait a couple of minutes for the reading to stabilize.

Note

: The electrode should be submerged approximately 4 cm

(1½”) in the solution.

• Adjust the right hand trimmer (

pH 7

) with the calibration

screwdriver until the LCD shows “pH 7.0”.

• Rinse and immerse the pH electrode in the pH 10.0 (or

pH 4.0) buffer and stir gently.

• Wait a couple of minutes and then adjust the left hand

trimmer (

pH 4

) until the LCD shows the value of the second

buffer.

The pH calibration is now complete.

pH CALIBRATION

HI 2114P/2

Double-junction, plastic body pH electrode with
2 m (6.6’) cable and BNC connector

HI 3214P/2

Double-junction, plastic body ORP electrode
with 2 m (6.6’) cable and BNC connector

HI 1283 *

Stainless steel grounding bar  with 2 m (6.6’)
cable

HI 70004P

pH 4.01 solution, 20 mL sachet (25 pcs)

HI 70007P

pH 7.01 solution, 20 mL sachet (25 pcs)

HI 70010P

pH 10.01 solution, 20 mL sachet (25 pcs)

HI 7004M

pH 4.01 solution, 230 mL bottle

HI 7007M

pH 7.01 solution, 230 mL bottle

HI 7010M

pH 10.01 solution, 230 mL bottle

HI 7020M

ORP 200/275 mV solution, 230 mL

HI 7092M

Oxidizing solution, 230 mL

HI 7061M

Electrode cleaning solution, 230 mL bottle

HI 710005

12 VDC power adapter, US plug

HI 710006

12 VDC power adapter, European plug

HI 710012

12 VDC power adapter, Australian plug

HI 710013

12 VDC power adapter, Southern Africa plug

HI 710014

12 VDC power adapter, UK plug

To be replaced only by authorized service technicians