background image

Instruction Manual

(HI 981402)

On-line, Waterproof

pH meter with Alarm

w w w . h a n n a i n s t . c o m

Hanna Instruments reserves the right to modify the design, construction

and appearance of its products without advance notice.

WARRANTY

HI 981402 is warranted for two years 

against defects in work-

manship and materials when used for their intended purpose and
maintained according to instructions.

The electrode is warranted for a period of six months.

This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered by
the warranty, you will be notified of the charges incurred. If the instrument
is to be returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department and then
send it with shipment costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.

All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.

Dear Customer,
Thank you for choosing a Hanna product.
This manual will provide you with the necessary information
for correct operation of the meter. Please read it carefully
before using the meter.
If you need additional technical information, do not hesitate
to e-mail us at

 [email protected]

.

This instrument is in compliance with the 

 directives.

Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
The meter is supplied with:

HI 1286

 pH electrode

• Calibration screwdriver
• 12 Vdc power adapter and instructions.

Note:

Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.

PRELIMINARY EXAMINATION

HI 981402 

is a pH meter specially designed to meet the

needs of simple continuous monitoring of pH.
The housing has been completely sealed against vapors and
humidity with IP54 rating.
Simply hang the meter right above the sample to be tested
for continuous measurement.
The 

HI 1286

 gel-filled pH electrode is interchangeable and

the BNC connector is protected behind a waterproof sheath.
The unique design of the electrode provides longer life even
in aggressive solutions.
This instruemnt also provides for a selectable setpoint and
alerts the user with a flashing LED alarm when an abnor-
mal situation occurs.
Measurements are highly accurate and the meter can be
calibrated at one or two points.
You no longer need to worry about battery changes either:
the unit runs without interruption on 12 Vdc power supply.

GENERAL DESCRIPTION

FUNCTIONAL DESCRIPTION

1. Molded eye
2. Alarm LED
3. “pH 4” calibration trimmer
4. BNC connector
5.

HI 1286 

pH electrode

6. 12 VDC power adapter
7. Power supply connector
8. Protective sheath
9. “pH 7” calibration trimmer

10. Liquid Crystal Display

pH ELECTRODE CONNECTION & MAINTENANCE

In order to protect the instrument against
vapors and humidity, the BNC connector is
shielded behind a waterproof sheath.

• Slide the protective sheath down. Connect the pH

electrode to the BNC connector and then slide the
protective sheath back up. Since the protective sheath is
rubberized to ensure maximum waterproof protection,
make sure the connector is completely covered.

• Do not be alarmed if white crystals appear around the

electrode protective cap. This is normal with pH elec-
trodes and they dissolve when rinsed with water.

SPECIFICATIONS

Range

0.0 to 14.0 pH

Resolution

0.1 pH

Accuracy 

(@ 25°C/77°F)

±0.2 pH

Typical EMC Deviation

±0.2 pH

Calibration

Manual, through offset and slope trimmers

pH Electrode

HI 1286

 (included)

Setpoint

3.0 to 11.0 pH

Hysteresis

±0.5 pH around setpoint

Alarm

LED blinks when pH is outside hysteresis range

Casing

IP54

Power supply

12 Vdc adapter (included)

Dimensions

86 x 110 x 43 mm (3.4 x 4.3 x 1.7’’)

Weight

150 g (5.3 oz.)

OPERATIONAL GUIDE

Reviews: