9

For greatest accuracy, frequent calibration of the instrument is recom-
mended. The instrument should be recalibrated for pH:

a)  Whenever the electrode is replaced.
b)  At least once a month.
c)  After testing aggressive chemicals.
d)  Where extreme accuracy is required.
e)  Whenever the CAL CHECK test fails (

HI9813-6

 only).

PREPARATION

Pour small quantities of pH7.01 (

HI7007

) or pH4.01 (

HI7004

) or 

pH10.01 (

HI7010

) solution into a clean beaker. 

To obtain accurate readings, use pH7.01 (

HI7007

) if you are going 

to measure neutral or close to neutral samples, pH4.01 (

HI7004

) if 

you are going to measure acidic samples or pH10.01 (

HI7010

) for 

alkaline measurements. 
If you need to calibrate to NIST standards, use pH6.86 (

HI7006

) instead 

of pH7.01 and pH9.18 (

HI7009

) instead of pH10.01.

PROCEDURE

•  Connect the probe and switch the meter on, then press the 

pH

 key 

to display pH measurement.

  Remove the protective cap from the probe, rinse and immerse it in 

the buffer and stir gently. Wait a couple of minutes for the reading 
to stabilize.

•  Take the temperature of the buffer solution, e.g. 

10.0 °C,  as follow: select the °C mode and read 
the displayed value.

•  Adjust the pH calibration knob until 

the LCD shows the pH value at the 
above temperature (see the pH versus 
temperature chart).

•  The pH calibration is now complete.

Notes: 

• The probe should be submerged approximately 4 cm (1½”) 

into the solution. The thermometer has to be located close 
to the probe.

 

•  If turning the knob the needed value can not be reached, 

clean the probe (see the “Probe Maintenance” section). If 
also after the probe cleaning the value can not be reached, 
replace the probe.

pH CALIBRATION

Summary of Contents for HI 9813-5

Page 1: ...1 Instruction Manual HI9813 5 HI9813 6 Portable pH EC TDS C Meters www hannainst com...

Page 2: ...5 and HI9813 6 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions Electrodes and probes are guarant...

Page 3: ...emove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment If noticeable damage is found please contact your local Hanna Instrumen...

Page 4: ...Liquid Crystal Display 3 pH range selection key 4 mS cm EC range selection key 5 ppm TDS selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 On Off key...

Page 5: ...ystal Display 3 pH range selection key 4 mS cm EC range selection key 5 ppm TDS range selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 Check key 10...

Page 6: ...onversion Factor variable 0 56 to 0 78 1 S cm according with TDS442 curve pH Calibration Manual 1 point through offset trimmer EC TDS Calibration Manual 1 point through slope trimmer EC TDS Tempera tu...

Page 7: ...0300 storage solution for 30 minutes to reactivate it To take a pH measurement simply submerge the tip 4 cm 1 of the probe into the sample to be tested Select the pH mode Stir briefly and wait a coupl...

Page 8: ...S mode TAKING TEMPERATURE MEASUREMENTS Submerge the tip 4 cm 1 of the probe into the sample to be tested Select the C mode Stir briefly and wait a couple of minutes for the reading to adjust and stabi...

Page 9: ...tead of pH7 01 and pH9 18 HI7009 instead of pH10 01 PROCEDURE Connect the probe and switch the meter on then press the pH key to display pH measurement Remove the protective cap from the probe rinse a...

Page 10: ...C F 4 01 6 86 7 01 9 18 10 01 0 32 4 01 6 98 7 13 9 46 10 32 5 41 4 00 6 95 7 10 9 39 10 24 10 50 4 00 6 92 7 07 9 33 10 18 15 59 4 00 6 90 7 05 9 27 10 12 20 68 4 00 6 88 7 03 9 22 10 06 25 77 4 01...

Page 11: ...DURE Pour approximately 4cm 1 of a conductivity calibration solution e g HI7031 into a beaker If possible use plastic beaker to minimize any EMC interference Immerse the probe in the solution Wait for...

Page 12: ...azardous area using a 9V alkaline battery Unscrew the three screws on the rear of the meter remove the battery compartment cover and replace the 9V battery with a new one while observing its polarity...

Page 13: ...0 the probe is broken and has to be replaced Press the Check key If the meter is calibrated Probe is OK message is displayed If cleaning is needed Clean Probe and Calibrate message is displayed 1 imm...

Page 14: ...I7009L pH9 18 buffer solution 500 mL bottle HI7010L pH10 01 buffer solution 500 mL bottle CONDUCTIVITY TDS CALIBRATION SOLUTIONS HI7031L 1413 S cm solution 500 mL bottle HI7442L 1500 ppm mg L solution...

Page 15: ...TIONS FOR USERS Before using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the suppli...

Page 16: ...ark 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Phone 800 426 6287 Fax 401 765...

Reviews: