Functional Description

Preliminary Examination

Remove the instrument from the packing material and examine 
it carefully to make sure that no damage has occurred during 
shipment. If noticeable damage is evident, contact your local Hanna 
Instruments Office.

Note: Save all packing material until you are sure that the 

instrument functions correctly. All defective items must be 

returned in the original packing together with the supplied 

accessories

.

Each meter is supplied with:

•  HI73127 pH electrode
•  HI73128 electrode removal tool
•  1.5V batteries (4 pcs.)
•  Instruction manual
•  Calibration card

General Description

HI98127

 (pHep®4) and 

HI98128

 (pHep®5) are waterproof 

pH and temperature meters. The housing has been completely 

sealed against humidity and designed to float. All pH readings are 

automatically temperature compensated (ATC), and temperature 

values can be displayed in °C or °F units.
These meters can be calibrated at one or two points with auto-

buffer recognition and against five memorized buffer values. A 

stability indicator is displayed on the LCD to ensure measurements 

are highly accurate.
At start-up, these meters indicate battery level and will display a 

low battery symbol to alert the user if the batteries need to be 

replaced. In addition, the Battery Error Prevention System (BEPS) 

avoids erroneous readings by turning the meter off when battery 

voltage levels are too low.
The included 

HI73127

 pH electrode can be easily replaced and the 

stainless steel encapsulated temperature sensor facilitates faster 

and more accurate temperature measurement and compensation.

1. ATC (Automatic Temperature Compensation) indicator
2. Stability indicator
3. Battery life percentage indicator
4. Low battery indicator
5. Secondary display
6. Primary display
7. Measuring unit for primary display

Specifications

Range

-2.0 to 16.0 pH (

HI98127

-2.00 to 16.00 pH (

HI98128

-5.0 to 60.0 °C / 
23.0 to 140.0 °F

Resolution

0.1 pH (

HI98127

0.01 pH (

HI98128

0.1 °C / 0.1 °F

Accuracy 

(@25 °C / 77 °F)

±0.1 pH (

HI98127

±0.05 pH (

HI98128

±0.5 °C / ±1.0 °F

Temperature 
Compensation

automatic

Environment

-5 to 50 °C (23 to 122 °F); RH 100%

Calibration

1 or 2 points with 2 sets of memorized 
buffers (pH 4.01 / 7.01 / 10.01 or 
pH 4.01 / 6.86 / 9.18)

Electrode

HI73127

 pH electrode (included)

Battery Type

1.5V (4 pcs.)

Battery Life

with BEPS / approx. 300 hours

Auto-off

after 8 minutes of non-use

Dimensions

171 x 41 x 26 mm (6.7 x 1.6 x 1.0”)

Weight

84 g (3.0 oz)

Calibration

For better accuracy, frequent calibration of the instrument is 
recommended. In addition, the instrument must be recalibrated 
whenever:

a) The pH electrode is replaced.
b) After testing aggressive chemicals.
c) Where high accuracy is required.
d) At least once a month.

Calibration procedure

From normal measuring mode, press and hold the  /MODE button 
until “OFF” on the secondary LCD is replaced by “CAL”. Release 
the button. The LCD enters the calibration mode displaying “pH 
7.01 USE” (or “pH 6.86 USE” if the NIST buffer set was selected). 
After 1 second the meter activates the automatic buffer recognition 
feature. If a valid buffer is detected then its value is shown on 
the primary display and “REC” appears on the secondary LCD. If 
no valid buffer is detected, the meter keeps the “USE” indication 
active for 12 seconds, and then it replaces it with “WRNG”, 
indicating the sample being measured is not a valid buffer.

To quit calibration and reset to default values
• After entering the calibration mode and before the first point is 

accepted, it is possible to quit the procedure and return to the last 
calibration data by pressing the  /MODE button. The secondary LCD 
displays “ESC” for 1 second and the meter returns to the normal 
measuring mode.

• To reset to the default values and clear a previous calibration, 

press the SET/HOLD button after entering the calibration mode 
and before the first point is accepted. The secondary LCD displays 
“CLR” for 1 second, the meter resets to the default calibration 
and the “CAL” tag on the LCD disappears.

LCD Description

Battery 

compartment

Liquid Crystal 

Display

/MODE 

button

SET/HOLD 

button

Temperature 

sensor

HI73127 pH 

electrode

1

2
3

4

7

6

5

All rights are reserved. Reproduction in whole or in part is prohibited without the 

written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode 

Island, 02895, USA.

Hanna Instruments reserves the right to m odify the design, construction, or 

appearance of its products without advance notice.

Recommendations for Users

For a 

single-point calibration

 with buffers pH 4.01, 9.18 or 10.01, 

the meter automatically accepts the calibration when the reading 
is stable; the meter displays the accepted buffer, with the message 
“OK 1”. After 1 second the meter automatically returns to the 
normal measuring mode. If a single-point calibration with buffer 
pH 7.01 (or pH 6.86) is desired, then after the calibration point 
has been accepted the  /MODE button must be pressed in order to 
return to normal mode. After the button is pressed, the meter shows 
“7.01” (or “6.86”) - “OK 1” and, after 1 second, it automatically 
returns to the normal measuring mode.

Note: A two-point calibration is recommended for better 

accuracy.

For a 

two-point calibration

, place the electrode in pH 7.01 (or pH 

6.86) buffer. After the first calibration point has been accepted, 
the “pH 4.01 USE” message appears. The message is held for 
12 seconds, unless a valid buffer is recognized. If no valid buffer is 
recognized, then the “WRNG” message is shown.
If a valid buffer (pH 4.01, 9.18 or 10.01) is detected, then the meter 
completes the calibration procedure. When the buffer is accepted, 
the LCD shows the accepted value with the “OK 2” message, and 
then the meter returns to the normal measuring mode.

Note: When the calibration procedure is completed, the “CAL” 

tag is turned on.

Before using Hanna Instruments products, make sure that they 
are entirely suitable for your specific application and for the 
environment in which they are used. Any variation intro-
duced by the user to the supplied equipment may degrade the  
meter' s performance. Avoid touching the electrode area. To avoid 
damages or burns, do not put the meter in microwave oven. For 
yours and the meter's safety do not use or store the meter in 
hazardous environments.

Summary of Contents for HI 98127

Page 1: ...INSTRUCTION MANUAL HI98127 HI98128 Waterproof pH Testers with Replaceable Electrode ...

Page 2: ...the meter has been calibrated CAL tag present on the LCD If measurements are taken in different samples successively rinse the probe thoroughly to eliminate cross contamination and after cleaning rinse the probe with some of the sample to be measured Setup Setup mode allows the selection of temperature unit and pH buffer set To enter the Setup mode press the MODE button until CAL on the secondary ...

Page 3: ...lacement The meter displays the remaining battery percentage every time it is switched on When the battery level is below 5 the symbol on the bottom left of the LCD lights up to indicate a low battery condition The batteries should be replaced soon If the battery level is low enough to cause erroneous readings the meter shows 0 BATT and the Battery Error Prevention System BEPS will automatically t...

Page 4: ...ectrode removal tool 1 5V batteries 4 pcs Instruction manual Calibration card General Description HI98127 pHep 4 and HI98128 pHep 5 are waterproof pH and temperature meters The housing has been completely sealed against humidity and designed to float All pH readings are automatically temperature compensated ATC and temperature values can be displayed in C or F units These meters can be calibrated ...

Page 5: ...escription Battery compartment Liquid Crystal Display MODE button SET HOLD button Temperature sensor HI73127 pH electrode 1 2 3 4 7 6 5 All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Hanna Instruments reserves the right to m odify the design construction or appearance ...

Page 6: ... of the instrument is recommended In addition the instrument must be recalibrated whenever a The pH electrode is replaced b After testing aggressive chemicals c Where high accuracy is required d At least once a month Calibration procedure From normal measuring mode press and hold the MODE button until OFF on the secondary LCD is replaced by CAL Release the button The LCD enters the calibration mod...

Page 7: ... to return to normal mode After the button is pressed the meter shows 7 01 or 6 86 OK 1 and after 1 second it automatically returns to the normal measuring mode Note A two point calibration is recommended for better accuracy For a two point calibration place the electrode in pH 7 01 or pH 6 86 buffer After the first calibration point has been accepted the pH 4 01 USE message appears The message is...

Page 8: ...d Goods Authorization RGA number from the Technical Service department and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection Electrode Code Description HI73127 Replaceable pH electrode HI73128 Electrode removal tool pH Buffer Solution Code Description HI70004P pH 4 01 solution 20 mL sachet 25 pcs HI70006P pH 6 86 solutio...

Reviews: