INSTRUCTION MANUAL

IST98127 06/18

HI98127 • HI98128

Waterproof pH Testers with Replaceable 

Electrode

Warranty

HI98127

 and 

HI98128

 are warranted for a period of one year 

against defects in workmanship and materials when used for their 
intended purpose and maintained according to instructions. The 
electrode is warranted for a period of six months. This warranty 
is limited to repair or replacement free of charge. Damage due to 
accidents, misuse, tampering or lack of prescribed maintenance 
is not covered. If service is required, contact your local Hanna 
Instruments Office. If under warranty, report the model number, 
date of purchase, serial number and the nature of the problem. 
If the repair is not covered by the warranty, you will be notified of 
the charges incurred. If the instrument is to be returned to Hanna 
Instruments, first obtain a Returned Goods Authorization (RGA) 
number from the Technical Service department and then send it 
with shipping costs prepaid. When shipping any instrument, make 
sure it is properly packaged for complete protection.

pH Electrode Maintenance

•  When not in use, rinse the electrode with water to minimize 

contamination and store it with a few drops of HI70300 storage 
solution in the protective cap. 
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE 
PURPOSES.

•  If the electrode has been left dry, soak in storage solution for 

at least one hour to reactivate it.

•  To prolong the life of the pH electrode, it is recommended 

to clean it monthly by immersing it in the HI7061 cleaning 
solution for half an hour. Afterwards, rinse it thoroughly with 
tap water and recalibrate the meter.

The pH electrode can be easily replaced by using the supplied tool 
(HI73128). 

Rotate the electrode 

counterclockwise. 

Pull the electrode out by using the other side of the tool. Insert 

a new pH electrode following the above instructions in reverse 

order.

To turn the meter on and check the battery status

Press and hold the  /MODE button until the LCD lights up. All the 
used segments on the LCD will be visible for 1 second (or as long 
as the button is pressed), followed by the percent indication of the 
remaining battery life (E.g. “% 100 BATT”).

Operational Guide

Taking measurements

Submerge the electrode in the solution 
to be tested while stirring it gently. The 
measurements should be taken when the 
stability symbol   on the top left of the 
LCD disappears. The pH value automatically compensated for 
temperature is shown on the primary LCD while the secondary LCD 
shows the temperature of the sample.

To freeze the display

While in measurement mode, press the 
SET/HOLD button. “HOLD“ appears on the 
secondary display and the reading will be 
frozen on the LCD (E.g. “pH 5.78 HOLD”). Press 
any button to return to normal mode.

Notes: Before taking any measurements make sure the 

meter has been calibrated (“CAL” tag present on the LCD).

If measurements are taken in different samples successively, 

rinse the probe thoroughly to eliminate cross-contamination and 

after cleaning, rinse the probe with some of the sample to be 

measured.

Setup

Setup mode allows the selection of temperature unit and pH buffer 
set. To enter the Setup mode, press the /MODE button until “CAL” 
on the secondary display is replaced by “TEMP” and the current 
temperature unit (E.g. “TEMP °C”). Then:

Electrode

Code 

Description

HI73127

Replaceable pH electrode

HI73128

Electrode removal tool

pH Buffer Solution

Code

Description

HI70004P

pH 4.01 solution, 20 mL sachet (25 pcs.)

HI70006P

pH 6.86 solution, 20 mL sachet (25 pcs.)

HI70007P

pH 7.01 solution, 20 mL sachet (25 pcs.)

HI70009P

pH 9.18 solution, 20 mL sachet (25 pcs.)

HI70010P

pH 10.01 solution, 20 mL sachet (25 pcs.)

HI770710P

pH 10.01 & 7.01 solution, 20 mL sachets 
(10 pcs., 5 ea.)

HI77400P

pH 4.01 & 7.01 solutions, 20 mL sachet  
(10 pcs., 5 ea.)

Electrode Cleaning Solution

Code

Description

HI7061M

Electrode cleaning solution, 230 mL bottle

Electrode Storage Solution

Code

Description

HI70300M

Electrode storage solution, 230 mL bottle

Other Accessories

Code

Description

HI740026P

Replacement 1.5V batteries (12 pcs.)

Accessories

Battery Replacement

The meter displays the remaining battery percentage 
every time it is switched on. When the battery level is below 5%, 
the   symbol on the bottom left of the LCD lights up to indicate a 
low battery condition. The batteries should be replaced soon. If the 
battery level is low enough to cause erroneous readings, the meter 
shows “% 0 BATT” and the Battery Error Prevention System (BEPS) 
will automatically turn the meter off. To change the batteries, remove 
the 4 screws located on the top of the meter (fig. 1).

Once the top has been removed, carefully replace the 

 

4 batteries located in the compartment while paying 

attention to their polarity (fig. 2).

Replace the top, making sure that the gasket is properly seated in 
place, and tighten the screws to ensure a watertight seal.

Note: Batteries should only be replaced in a safe area using the 

battery type specified in this instruction manual. Old batteries 

should be disposed in accordance with local regulations. 

•  for °C / °F selection: Use the SET/HOLD button. After the 

temperature unit has been selected, press the  /MODE button 
to enter the buffer set selection mode; press the  /MODE 
button twice to return to the normal measuring mode.

•  to change the calibration buffer set: After setting the 

temperature unit, the meter will show the current buffer set: 
“pH 7.01 BUFF” (for pH 4.01 / 7.01 / 10.01) or “pH 6.86 
BUFF” (for NIST 4.01 / 6.86 / 9.18). Change the set with the 
SET/HOLD button, then press  /MODE to return to normal 
measuring mode.

fig. 1

fig. 2

To turn the meter off

While in normal mode, press the  /MODE button. “OFF” will 
appear on the secondary display. Release the button.

^

Insert the tool into the electrode 

cavity as shown above.

Summary of Contents for HI 98127

Page 1: ...INSTRUCTION MANUAL HI98127 HI98128 Waterproof pH Testers with Replaceable Electrode ...

Page 2: ...the meter has been calibrated CAL tag present on the LCD If measurements are taken in different samples successively rinse the probe thoroughly to eliminate cross contamination and after cleaning rinse the probe with some of the sample to be measured Setup Setup mode allows the selection of temperature unit and pH buffer set To enter the Setup mode press the MODE button until CAL on the secondary ...

Page 3: ...lacement The meter displays the remaining battery percentage every time it is switched on When the battery level is below 5 the symbol on the bottom left of the LCD lights up to indicate a low battery condition The batteries should be replaced soon If the battery level is low enough to cause erroneous readings the meter shows 0 BATT and the Battery Error Prevention System BEPS will automatically t...

Page 4: ...ectrode removal tool 1 5V batteries 4 pcs Instruction manual Calibration card General Description HI98127 pHep 4 and HI98128 pHep 5 are waterproof pH and temperature meters The housing has been completely sealed against humidity and designed to float All pH readings are automatically temperature compensated ATC and temperature values can be displayed in C or F units These meters can be calibrated ...

Page 5: ...escription Battery compartment Liquid Crystal Display MODE button SET HOLD button Temperature sensor HI73127 pH electrode 1 2 3 4 7 6 5 All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Hanna Instruments reserves the right to m odify the design construction or appearance ...

Page 6: ... of the instrument is recommended In addition the instrument must be recalibrated whenever a The pH electrode is replaced b After testing aggressive chemicals c Where high accuracy is required d At least once a month Calibration procedure From normal measuring mode press and hold the MODE button until OFF on the secondary LCD is replaced by CAL Release the button The LCD enters the calibration mod...

Page 7: ... to return to normal mode After the button is pressed the meter shows 7 01 or 6 86 OK 1 and after 1 second it automatically returns to the normal measuring mode Note A two point calibration is recommended for better accuracy For a two point calibration place the electrode in pH 7 01 or pH 6 86 buffer After the first calibration point has been accepted the pH 4 01 USE message appears The message is...

Page 8: ...d Goods Authorization RGA number from the Technical Service department and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection Electrode Code Description HI73127 Replaceable pH electrode HI73128 Electrode removal tool pH Buffer Solution Code Description HI70004P pH 4 01 solution 20 mL sachet 25 pcs HI70006P pH 6 86 solutio...