HI98103 

 

pH Tester

INSTRUCTION MANUAL

Preliminary Examination

Specifications

Range

0.0 to 14.0 pH

Resolution

0.1 pH

Accuracy 

(@25 °C/77 °F)

±0.2 pH

Calibration

automatic, one or two-point

Electrode

HI1271

 (included)

Battery Type

CR2032 Li-ion

Battery Life

approximately 1000 hours of continuous use

Auto-off

8 minutes, 60 minutes or can be disabled

Environment

0 to 50 °C (32 to 122 °F); RH 95% max

Dimensions

50 x 174 x 21 mm (2 x 6.8 x 0.9“) 

Weight

50 g (1.8 oz.)

Liquid Crystal 
Display

ON/OFF 
Button

pH Probe

Maximum 
Immersion 
Level

Recommended 
Immersion 
Level 4 cm (1.5”)

Operation

Press the ON/OFF button to turn the 
meter on. All tags will be displayed.

The meter will go into measurement 
mode: current reading and 
calibrated buffers are displayed.

Thank You

Thank you for choosing a Hanna Instruments product. Please read 
this instruction manual carefully before using the instrument. 

For more information about Hanna Instruments and our products, visit  
www.hannainst.com.

For technical support, contact your local Hanna Instruments Office 
or email us at [email protected]

Find your local Hanna Instruments Office on www.hannainst.com

Remove the meter from the packing material and examine it carefully to 
make sure that no damage has occurred during shipment. If noticeable 
damage is evident, contact your local Hanna Instruments Office.
Each meter is supplied with:

• pH 4.01 Buffer Solution Liquid Sachet (2 pcs.)
• pH 7.01 Buffer Solution Liquid Sachet (2 pcs.)
• General purpose cleaning solution (2 pcs.)
• Instruction manual
• Quality Certificate

Note: Save all packing material until you are sure that the instrument 

functions correctly. All defective items must be returned in the original 

packaging with the supplied accessories.

Preparation:

The pH electrode is shipped dry. Before using the Checker

®

, remove the protective 

cap and condition the electrode by soaking the tip (bottom 4 cm (1.5”)) in  
pH 7.01 buffer solution for several hours. Then follow the calibration procedure.

•  Do not be alarmed if white crystals appear around the cap. This is 

normal with pH electrodes and they dissolve when rinsed with water.

•  Connect the electrode to the meter.
•  Turn the Checker

®

 on by pressing ON/OFF button.

•  Remove the protective cap and immerse the tip of the electrode in the 

sample to be tested.

•  Stir gently and wait for a stable reading.

NEVER IMMERSE THE ELECTRODE OVER THE MAXIMUM IMMERSION LEVEL. 
THE CONNECTOR MUST ALWAYS BE CLEAN AND DRY.

•  After use, rinse the electrode with water and store it with a few drops of  

HI70300 storage solution in the protective cap.

•  Replace protective cap after each use.

DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.

Meter Overview

The Checker

®

 features a low 

battery indicator. When the battery 
is running low, the tag will blink 
on screen. When the battery has 
been depleted, “Erb” will appear on 
screen and the meter will turn off.

Place meter in calibration mode. 
Press and hold ON/OFF until “CLr” 
is displayed. The meter will now be 
at default calibration. No tags will 
be shown in measurement mode 
until calibration is performed.

In calibration mode, if the meter 
displays an “Err” message 
when in the correct fresh buffer 
solution then the probe should be 
cleaned.  Place the probe in the

 

HI700601 

cleaning solution for 

15 minutes. Rinse with purified 
water and place in storage solution 
for 1 hour before calibrating. If the 
“Err” message persists then the 

HI1271

 probe should be replaced.

Clear Calibration

“Err” Message

Battery Indicator

From measurement mode, press and 
hold the ON/OFF button. The meter 
will cycle through “OFF,” “CAL,” 
then current auto-off setting.

The default setting is 8 minutes 
(“d08”). Press ON/OFF button to 
change. “d60” is auto-off after 
60 minutes, and “d--” disables 
the auto-off feature. Press and 
hold the button to exit the menu.

Auto-off

Summary of Contents for HI 98103

Page 1: ...HI98103 pH Tester INSTRUCTION MANUAL ...

Page 2: ...re that the instrument functions correctly All defective items must be returned in the original packaging with the supplied accessories Preparation ThepHelectrodeisshippeddry BeforeusingtheChecker removetheprotective cap and condition the electrode by soaking the tip bottom 4 cm 1 5 in pH 7 01 buffer solution for several hours Then follow the calibration procedure Do not be alarmed if white crysta...

Page 3: ...o off 8 minutes 60 minutes or can be disabled Environment 0 to 50 C 32 to 122 F RH 95 max Dimensions 50 x 174 x 21 mm 2 x 6 8 x 0 9 Weight 50 g 1 8 oz Liquid Crystal Display ON OFF Button pH Probe Maximum Immersion Level Recommended Immersion Level 4 cm 1 5 Operation Press the ON OFF button to turn the meter on All tags will be displayed The meter will go into measurement mode current reading and ...

Page 4: ...en in the correct fresh buffer solution then the probe should be cleaned Place the probe in the HI700601 cleaning solution for 15 minutes Rinse with purified water and place in storage solution for 1 hour before calibrating If the Err message persists then the HI1271 probe should be replaced Clear Calibration Err Message Battery Indicator From measurement mode press and hold the ON OFF button The ...

Page 5: ... pH 4 01 or pH 10 01 B One Point Meter will exit to measurement mode and the calibration tag will be displayed pH 4 01 will then blink on the display Two Point Use pH 4 01 or pH 10 01 to perform a two point calibration The value is automatically recognized and displayed with the blinking stability indicator When the reading is stable the stability indicator will disappear Sto will be displayed whe...

Page 6: ...meter is not in use it is important to add several drops of storage solution to the protective cap to keep the probe hydrated If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Forimprovedaccuracyitisrecommendedtocalibratetoaminimumoftwo points It is important to use buffers that bracket the expected value of the sample to be tested For example if the expected value is pH 8...

Page 7: ...department and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA IST98103N 06 18 Electrode Storage Solution Code Description HI70300L electrode ...