20
21
It is possible to delete the calibration and
restore factory calibration.
• Press
GLP/
to enter
GLP mode.
To display the calibration date:
• Press
GLP/
to enter
GLP mode. The
calibration month and day will appear on
the main display and the year on the
secondary display.
• Press
READ/ /TIMER
to enter in the
factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.
• Press
ZERO/CFM
to restore the factory
calibration or press
GLP/
again to
abort factory calibration restore.
• The instrument briefly notifies
“done”
when restores factory calibration and
returns to
measurement mode.
BATTERY MANAGEMENT
To save battery, the instrument shuts down after 10 minutes of non-use in
measurement mode
and after 1 hour of non-use in
calibration mode.
If a valid measurement was displayed before
auto-shut off, the value is displayed when
the instrument is switched on. The blinking
“ZERO”
means that a new zero has to be
performed.
One fresh battery lasts for around 750 measurements, depending on the light level.
The remaining battery capacity is evaluated at the instrument startup and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken anymore, the instrument
shows “
dead batt
” and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
In the GLP mode, the last user calibration date can be consulted and the factory calibration can
be restored.
LAST CALIBRATION DATE
• If no calibration was performed, the
factory calibration message,
“F.CAL”
will
appear on the main display and the
instrument returns to
measurement mode
after three seconds.
FACTORY CALIBRATION RESTORE
G L P
• The instrument will return automatically
to
measurement mode by displaying
dashes on the LCD.