• After a few seconds, the display will show “
-0.0-
”. The meter is now zeroed and ready for
measurement.
• Remove the cuvette.
• Add the content of one packet of
HI93750B‑0
powder reagent. Replace the cap and shake
gently 1 minute.
• Press and hold
READ /TIMER
for three seconds. The display will show the countdown prior to
measurement. The beeper will signal a tone at the end of countdown period to alert the user.
Alternatively, wait for 3 minutes.
• Invert the cuvette for 5 times, replace the cuvette into the cuvette holder and ensure that the
notch on the cap is positioned securely into the groove. Press
READ /TIMER.
In both cases,
the lamp, cuvette and detector icons will appear on the display, depending on the measurement
phase.
• At the end of measurement, the instrument directly displays concentration in mg/L of potassium
on the LCD.
• If the analysis result is over the method range use P2 procedure.
HI9
37
50
B-0
MEASUREMENT PROCEDURE
3"
3’
or
x5